7

confirm and modify the parameters have been set. 

d. "AT" self-adjusting indicator: The instrument activates self-adjusting function keys; 

e. "OUT" master output indication: light during heating; 

f. "ALM" signal indicator; 

g.  "

"  "

"  Addition  and  Subtraction  Key:  When  setting  parameters,  it  is  used  to  increase  or 

decrease the value. 

2.Temperature setting: 

Press “

” key, when “SP” appears in “PV” window, press “

  ” and “

  ” to increase or 

decrease  according  to  the  required  temperature  value.  When  the  value  of  “

”  display  window 

reaches the required temperature value, ie After the setting is completed, press "

" to confirm. 

 

Method of operation 

a. Put the digestive tube containing samples in vitro digestion rack, will waste gas in the in 

vitro digestion absorption equipment rack (hands holding a waste gas absorption bottle, to absorb 

the gas bottle seal on the digestive tube, cover can.) 

b.  According  to  the  above  diagram,  connect  the  tap  water  and  the  pumping  pump,  and  pay 

attention  to  the  proper  water  flow  (3/4l  /min).  First  pick  good  extraction  tee  (the  suction  filter 

pump),  adjust  flow  after  connected  to  the  gas  absorption  tube  (besides  using  standard,  our 

company  also  with  another  high  matches:  vacuum  pumps,  filters,  and  exhaust  gas  absorption 

equipment connection. This method can neutralize the harmful gases such as SO2, which is easier 

to operate. 

c. Connect the power cord in the fitting, open the main switch of the ship type power supply, 

the temperature instrument is bright, and the first infrared quartz tube is energized to start the work; 

Press  the  power  switch  1,  and  the  second  infrared  quartz  tube  begins  to  heat;  Press  the  power 

switch 2 the third infrared quartz tube begins to heat. 

d. According to the instrument instructions and set the temperature in two stages: first of all, 

first set to a temperature of 200 degrees, after reaching the temperature continue to heat the 20-30 

min,  and  then  reset  the  temperature  continue  heating  30  to  60  min  after  450  degrees,  then  the 

sample  will  gradually  stop  boiling,  and  then  began  to  change  color,  when  the  sample  color 

becomes completely  transparent  blue-green liquid continue  heating 10-20  min, after  digestion to 

complete  at  this  time.(such  as  moisture  content  is  exorbitant,  heating  of  the  digestive  wall  with 

black  objects,  to  extend  the  heating  time  until  the  color  of  the  sample  becomes  transparent 

blue-green liquid) 

Generally,  the  time  required  is  60-90min  from  the  boot  to  the  sample,  which  is  determined 

according to the sample (the higher the nitrogen content, the longer the digestion time). 

e.  After  the  digestion  is  completed,  it  will  be  moved  to  the  test  tube  cooling  rack  together 

with the tube digester and exhaust gas absorption device. 

 

Содержание DW-KDN-08D

Страница 1: ...s DW KDN 08D Drawell International Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Страница 2: ...tures 5 2 Technical Parameters 5 Method of connection of digester furnace 6 Digestion furnace temperature control 6 1 Temperature control instrument button description 6 2 Temperature setting 7 Method of operation 7 Precautions 8 List of attachments 9 ...

Страница 3: ...he message please give the message and we express our heartfelt thanks for it Shanghai Drawell Scientific Instruments Co Ltd Tel 0086 21 50966080 Security warning Danger May constitute property damage or personnel injury please use as required SPECIAL POWER SOURCE GND FLAMMABLE AND EXPLOSIVE No RENOVATION a Please use the power supply and voltage conforming to the requirements of this product b Th...

Страница 4: ...vironment requires no dust inflammable and explosive gas h The working environment temperature is 10 45 degrees relative humidity is less than 80 i The instrument must be installed in a place without direct sunlight j The reagent and other solutions are corrosive and should be equipped with protective equipment and safe operation during operation ...

Страница 5: ...20ml c Capacity of digestive tube 300ml standard d Determination range 240mgN e Temperature control range at room temperature to 550 f Heating rate 30 degrees min g Temperature control accuracy 1 h Power consumption digester 8 holes 2000W i Power source 220 V 10 50 60HZ j Appearance size 435 270 225 mm k Weight 8 6kg In accordance with the principle of the kjeldahl method the apparatus is widely u...

Страница 6: ...h rubber hose connection The filter contains 500 1000ml of lye which is neutralized by harmful gases such as SO2 which can be used to filter the harmful substances of adsorption particles Digestion furnace temperature control D type digital digestion oven using a new generation of digital temperature controller to control the digester heating can display the target temperature and the actual tempe...

Страница 7: ...thod can neutralize the harmful gases such as SO2 which is easier to operate c Connect the power cord in the fitting open the main switch of the ship type power supply the temperature instrument is bright and the first infrared quartz tube is energized to start the work Press the power switch 1 and the second infrared quartz tube begins to heat Press the power switch 2 the third infrared quartz tu...

Страница 8: ...waste gas from overflowing when taking the exhaust gas absorption bottle first pour the bottle to one side and then take it off the whole so as to avoid the pressure to bring the digestive tube together with the accident c Since the location of the alimentary canal digestion process the surrounding and the temperature of the center position is a bit poor so some sample color change first appeared ...

Страница 9: ...8 holes 1 3 Digestive tube cooling rack 8 holes 1 4 Digestive tube Φ40 300mm 8 5 Exhaust absorption Φ50 280mm 2 6 Exhaust gas absorption bottle seal Φ55 28mm 8 7 Extraction tee Teflon 1 8 Power cable AC 220 1 9 Y type tee PC tube 1 10 Fuse tube 10A 2 11 6 11 Rubber tube Meter 4 12 Certification Sheet 1 ...

Страница 10: ...ational Technology Limited Shanghai Drawell Scientific Instrument Co Ltd Add Suite 1506 Lane581 XiuChuan Rd PuDong New Area Shanghai China Tel Fax 0086 21 50966080 Web www drawell com cn Email sales01 drawell com cn ...

Отзывы: