19
Fig 36 cubic fit
Fig 37 linear fit
3.3.5
Press
【
SAVE
】
to save calibration if required
3.3.6
Press
【
ESC/STOP
】
to exit
4. Quantitative Test
Before test,the standard curve must be obtained.There are three ways
for you to obtained it (a, b or c).
a) Standard curve built up and saved in the instrument.
In Fig 33 press
【
Load
】and then press【∧】
or
【∨】
to select the
file with type ***.fit. At last press
【
ENTER
】
TO comfirm.
b) Known standard curve, which is not saved in the instrument.
See 3.2. For Fig 29 enter a known standard curve directly.
c)
Use the standard samples for the test. First the standard curve must
be established using the method shown in 3.3.
Note: All sample results must be taken in screen Fig26.
4.1
Push the blank cuvette into the Reference Light Path and Main Light
Path and press
【
ZERO
】
for blanking.
4.2
Pull the sample cuvette into Main Light Path, press the key
【
START
】
,
the results will be displayed on the screen (Fig 38).