12
Fig 16
Fig 17
Before measurement
Make a blank reference solution by filling a clean cuvette (or test tube) half
full with distilled or de-ionized water or other specified solvent. Wipe the
cuvette with tissue to remove the fingerprints and droplets of liquid.
Analyze Sample
For different user’s requirements, we provide different test methods.
Basic Mode
Push the blank cuvette into the Reference Light Path and Main Light Path. In
main menu (Fig7),press
【
1
】
to enter “Basic mode” test . After automatically
blanking, it will display as Fig 18 (automatic changer installed) or Fig 19
( automatic changer uninstalled) and wait for the operator.
【
ESC/STOP
】
to
exit.
Note:
If no automatic changer installed “cell #1” and “Max E” will disappear
in Fig18
Fig18