background image

8     Setup and Programming / Software Flow Chart

GENERAL
Programming and viewing of the various setup and operating
parameters are accomplished using the front panel back lit,
two line, sixteen character wide LCD along with the LEFT,
RIGHT, UP, DOWN and ENTER buttons. Generally, the name
of the parameter is on the top line of the display and the
setting value is on the bottom line. For some parameters,
both the parameter name and value are on the bottom line.

To observe a certain parameter setting without intending to
change its value, just use the LEFT and RIGHT arrow
buttons to navigate through the menus shown in the software
flow chart below. Generally, the current setting for each
parameter is shown on the bottom line of the display. Note
that depending upon certain settings, some screens are not
needed and will be skipped.

To make a change in the displayed parameter  and its setting
and if this is the initial setup, you will want to enter the ‘adjust’
mode. To do this, press the ENTER button that is located in
the center of the four arrow buttons and hold in for several
seconds until a flashing 

^

 appears in the upper right hand

corner of the display.   After you are in the adjust mode, use
the LEFT and RIGHT arrow buttons to navigate among
screens and use the UP and DOWN arrows to change the
parameter setting.

When certain numerical values are to be adjusted, one or
more  asterisks (

*

) will appear above the digits to be ad-

justed.  Use the UP and DOWN keys to change the value

of the digit, and the LEFT or RIGHT keys to move to another
digit.  Note that the astrisk(s) will not appear when scrolling
through adjustable parameters using the LEFT arrow key
unless it has already been activated using the RIGHT arrow
key.

When ENTER is pressed, the new settings will be loaded
and stored and the unit will exit the ‘adjust’ mode. You may
wish to not press ENTER until you have gone through all
screens and settings and then press ENTER to save and
load all changes in one step, OR you can store just one or
several parameters at a time and reenter adjust mode to set
the next.  Either method is acceptable.

The OBM100 has displays that show the number of frames
per second that are input from the ASI inputs and are being
sent to the RF output.  Each frame that comes from the ASI
inputs contains seven packets of 188 bytes each (1316
bytes).  If a connection to the headend (as defined by the
destination address) has not been established through an
ARP reply, then the characters that display the number of
frames per second for the ASI input is replaced with a "NC"
display.

 The frames being sent to the RF output contain between one
and seven packets of 188 bytes, depending upon what is
being sent to the OBM100 from the headend controller.  The
headend controller will need to send frames with seven
packets in each frame to allow the unit to function properly
when high data rates are being sent through the RF output.

OBM100 QPSK DATA MODULATOR SOFTWARE FLOW CHART

SETTINGS
Frequency:  
Set the output frequency of the OBM100 as
required by the system map design.  The frequency range is
from 70 to 130 MHz, adjustable in .05 MHz steps.

Output:  Set the desired RF output signal level. The avail-
able range is from 30.0 to 52 dBmV in .5 dBmV steps.  The
output accuracy is 

± 1 dB.

OUTPUT MODE:  For normal operation, select NORMAL.
For system level set up, choose CW to provide a CW carrier
at the center frequency of the output channel for use in
leveling a system when a QAM power meter is not available.
In the CW mode, the CW carrier can be measured on a

spectrum analyzer without a need to apply a bandwidth
correction or it can be measured with an analog meter tuned
to channel center.  The CW power measured will equal the
channel QAM power when the modulator is returned to
normal output mode. Usually QAM signals are set 5 dB
below analog NTSC channels when balancing a system.

The PRBS menu entry is used for testing purposes only and
should normally not be selected by the user.

Symbol Rate:  This setting allows selection of the output
symbol rate.  Set as required by the set top boxes.  The
choices are .772, 1.024, and 1.544 MHz.

Freq  xxx.xx MHz

**

Output

Output
 Mode

Symbol Rate

70 - 130 MHz

NORMAL
CW
PRBS

.772

1.024
1.544

Dest MAC

FFFFFFFFFFFF

000000000000 - 

FFFFFFFFFFFF

Dest addr

255.255.255.255

*

*

Dest Port 65535

000.000.000.000 -

255.255.255.255

00000 -

65535

DHCP
enable

ON
OFF

255.255.255.255

*

IP addr

GATEWAY addr

255.255.255.255

*

SUBNET mask

255.255.255.255

*

255.255.255.255

000.000.000.000 -

255.255.255.255

000.000.000.000 -

255.255.255.255

000.000.000.000 -

DHCP ON (Adjust Mode)

DHCP OFF

(Adjust & Normal Modes)

DHCP ON (Normal Mode)

DHCP OFF

(Adjust Mode)

(Not User Settable)

30.0 - 52.0 dBmv

OBM100  1.4  c

DHCP...........

DHCP...........

or

xxx.xxx.xxx.xxx

Board Setup...

MAC xxxxxxxxxxxx 

to RF Output

xxx Frames/S

xxx Frames/S

from ASI Input

DHCP OFF

(Normal Mode)

Up & Down

buttons in

normal mode

only.

Содержание OBM100

Страница 1: ...tion Manual TM is a registered trademark of the R L Drake Company Copyright 2009 R L Drake LLC P N 3852496 01 2009 Printed in U S A OBM100 QPSK Data Modulator ENTER OBM100 QPSK DATA MODULATOR DHCP OBM...

Страница 2: ...be moved with care Quick stops excessive force and uneven surfaces may cause the product and cart combina tion to overturn The lightning flash with arrow head symbol within an equilateral triangle is...

Страница 3: ...ll fit into the power outlet only one way This is a safety feature If you are unable to insert the plug fully into the outlet try reversing the plug If the plug should still fail to fit contact your e...

Страница 4: ...acement Parts When replacement parts are required be sure the service technician has used replacement parts specified by the manufacturer or have the same characteristics as the original part Unauthor...

Страница 5: ...ay Accuracy Output Level Stability Frequency Stability Output Impedance Mode Symbol Rate FEC ASI INPUTS Four ASI Inputs Data Rate ETHERNET INPUT UDP to UDP Frame Spacing UDP to UDP Frame Spacing UDP t...

Страница 6: ...t INSTALLATIONANDMOUNTINGNOTES This equipment is designed to be installed in a standard 19 rack When the unit is mounted above or below other rack mounted equipment a 1U space 1 75 should be left betw...

Страница 7: ...er the adjust mode or to save and load a new setting or settings after adjustment Hold for approximately 2 seconds until a flashing appears in the upper right hand corner of the display to enter the a...

Страница 8: ...h 1316 bytes If a connection to the headend as defined by the destination address has not been established through an ARP reply then the characters that display the number of frames per second for the...

Страница 9: ...utomatically by the Headend Controller ADDITIONALINFORMATION The OBM100 display showing the version number can also display additional unit status and error conditions A c will appear to the right of...

Страница 10: ...ves at R L DRAKE if we have all the facts If we must call you it may take longer R L DRAKE is not responsible for damage caused by lightning nonprofes sional alterations acts of God shipping damage po...

Страница 11: ...are important to obtain proper performance from this product Therefore carefully read the Instruction Manual This warranty does not apply to any defect that R L DRAKE LLC determines is due to 1 Impro...

Страница 12: ...R L Drake LLC 230 Industrial Drive Franklin Ohio 45005 U S A Customer Service and Parts Telephone 1 937 746 6990 Telefax 1 937 806 1576 World Wide Web Site http www rldrake com...

Отзывы: