18
Instructions for use
|
Dräger X-node
en
|
Technical operating data
13.1
Technical data when operated with CO
2
sensor 6800482
13.2
Technical data ambient sensors
Interface:
USB interface
Housing
Housing material:
Polycarbonate
Housing protection rating:
IP65
Dimensions:
Approx. 170 mm x 90 mm
Weight:
Approx. 300 g
Ambient conditions
Temperature:
-20 °C to +50 °C
Humidity:
5 % to 95 % r.h.
Pressure:
700 hPa to 1300 hPa
Electrical safety:
Radio approval:
EN 62311:2008
EN 300 200-2:2016
EN300 328: 2019-07
Operating pressure range: 700 hPa to 1300 hPa
Measuring range:
0 to 10,000 ppm
Accuracy at 25 °C
±(30 ppm, +3 % of measured value)
Alarm threshold A1:
1000 ppm
Alarm threshold A2:
2000 ppm
Ambient pressure:
Accuracy 700 hPa to 1100 hPa = ±
1.0 hPa, 1100 hPa to 1300 hPa = ±
1.5 hPa
Temperature:
Accuracy 0 °C to 50 °C =
± 1.0 °C
Humidity:
5 % to 95 % r.h., accuracy ± 5 % r.h.