background image

PSS

® 

4000 Series

Self-contained breathing apparatus

Instructions for Use

i

Dow Corning

®

 and Molykote

®

 are registered trademarks of Dow Corning Corporation

3363923 (A3-D-P) Page 2 of 3

3. Press the rubber cover at the front of the lung demand valve to vent

the system fully. Press the reset button when using a positive-pressure
lung demand valve.

4. Release the waist belt buckle.
5. Lift the shoulder strap buckles to loosen the straps.
6. Remove the breathing apparatus and face mask.
7. Carry out the after use tasks in the maintenance table (see Section 5).
8. Remove the air cylinder (see Section 3.5.2) if required.
9. Pass the breathing apparatus to the service department with details of

any faults or damage that occurred during use.

3.5

Common user tasks

3.5.1

Visual inspection

A visual inspection must check the full breathing apparatus including all
component parts and accessories. Check that the equipment is clean and
undamaged, paying particular attention to pneumatic components, hoses
and connectors. Typical signs of damage that may affect the operation of
the breathing apparatus include impact, abrasion, cutting, corrosion and
discolouration. Report damage to service personnel and do not use the
apparatus until faults are rectified.

3.5.2

Air cylinder fitting and removing

WARNING

High-pressure air release may cause injury to the user or other
personnel near the breathing apparatus. Close the cylinder valve
and fully vent the system before attempting to disconnect an air
cylinder.
Impact damage to the cylinder valve or reducer connector may
prevent valve connection or cause an air leak. Handle the air
cylinder and breathing apparatus with care.

NOTICE

For other cylinder connector types, refer to the Instructions for Use
supplied for the connector.

Fitting a single cylinder

1. Check the threads of the cylinder valve port and the pressure reducer.

Ensure that the O-ring seal (Fig D, Item 1) and the sintered filter (2) in
the reducer are clean and undamaged.

2. Lay the apparatus horizontal, with the reducer uppermost, and fully

extend the cylinder strap.

Twin cylinder configuration: move the centre strap over the left-

hand side (Fig E, Item 1).

3. Insert the cylinder through the loop of the strap, and align the valve with

the reducer.

4. Lift the cylinder and space frame into the vertical position (supported

on the end of the cylinder opposite the valve).

5. Tighten the hand wheel of the reducer, using only the thumb and index

finger, until a definite metal-to-metal contact is felt. 

Do not

 use tools or

over tighten.

6. Place the unit back into the horizontal position.
7. Take up the slack in the cylinder strap (Fig F).

Twin cylinder configuration: if fitted, feed the cylinder strap through

the metal buckle (Fig G, Item 1).

8. Pull the strap over the cylinder to operate the cam lock (Fig H).
9. Secure the strap end:

Single cylinder configuration: using the hook-and-loop fastener.

Twin cylinder configuration: feed the cylinder strap through the

metal buckle of the centre strap (Fig I) and secure using the hook-

and-loop fastener.

Fitting twin cylinders

Fitting twin cylinders requires a PSS

® 

4000 variant capable of accepting

twin cylinders. Fitting twin 6 litre or 6.8 litre composite material cylinders
requires a compatible connector piece (see Section 2.2.1). Fitting twin
4 litre steel cylinders requires a compatible connector piece and slim hand
wheel (see Section 2.2.1).

1. Check the threads of the cylinder valve port and the pressure reducer.

Ensure that the O-ring seal (Fig D, Item 1) and the sintered filter (2) in
the reducer are clean and undamaged.

2. Lay the apparatus horizontal, with the reducer uppermost, and fully

extend the cylinder strap.

3. If not fitted, fit the connector piece on to the reducer (Fig J, T-piece

shown for illustrative purposes). Tighten the hand wheel of the reducer
using only the thumb and index finger until a definite metal-to-metal
contact is felt. Do not use tools or over tighten.

4. Insert the cylinder through the loop of the strap, and align the valve with

the connector piece.

5. Tighten the hand wheel of the connector piece as previously described

in Step 3 (Fig K).

6. If required, fit the anti-vibration straps on to the hand wheel.
7. Take up the slack in the cylinder strap (Fig L).
8. Pull the strap over the cylinder to operate the cam lock (Fig M).
9. Secure the strap end using the hook-and-loop fastener (Fig N).

(Release the cam lock to adjust the position of the cylinder strap if
necessary.)

Removing a single cylinder

1. Close the cylinder valve and fully vent the system.
2. Lay the apparatus horizontal, with the cylinder uppermost.
3. Unfasten the cylinder strap.
4. Lift the strap against the cam lock to release the buckle tension, and

then loosen the strap.

5. Disconnect the cylinder valve from the pressure reducer.
6. Lift the cylinder away from the reducer and remove the cylinder.

Removing twin cylinders

1. Close the cylinder valves and fully vent the system.
2. Lay the apparatus horizontal, with the cylinder uppermost.
3. Release the cam lock, and then loosen the strap.
4. Remove the anti-vibration straps, if fitted, and fully unscrew the hand

wheels of the connector piece.

5. Lift the cylinders away from the connector piece and remove the

cylinders.

6. Remove the connector piece from the reducer.

!

i

i

3.5.3

Functional testing

WARNING

Failure of the breathing apparatus to meet any of the standards or parameters described in the functional tests indicates a system fault. Report
the fault to trained service personnel or contact Dräger. Do not use the breathing apparatus until the fault condition is rectified.

Assemble the breathing apparatus as described in the preparation for use (see Section 3.1) before commencing any functional testing.

Leak test and whistle warning test

1. Press the reset button when using a positive-pressure lung demand valve.
2. Slowly and fully open the cylinder valve (anticlockwise). During pressurization a momentary sounding of the whistle will occur.
3. Fully close the cylinder valve.
4. After one minute, check the contents gauge and then reopen the cylinder valve. The gauge must not show an increase in pressure of more than

10 bar (one radial marking on the gauge face). Investigate and repair a failed leak test (see Section 4), and then repeat the leak test.

5. Fully close the cylinder valve.
6. Observe the contents gauge and slowly release the pressure as follows:

a. Cover the outlet port of the valve with the palm of the hand.
b. Press the rubber cover to switch on the positive pressure.
c. Vent the system by carefully lifting the palm of the hand from the outlet port to maintain a slow pressure decrease.

7. The whistle must begin to sound in the range 60 bar to 50 bar.
8. Continue to vent the system until fully exhausted.
9. Press the reset button when using a positive-pressure lung demand valve.

4

Troubleshooting

The troubleshooting guide shows fault diagnosis and repair information applicable to breathing apparatus users. Further troubleshooting and repair
information is available in Instructions for Use supplied with associated equipment.

Where the troubleshooting guide shows more than one fault or remedy, carry out repair actions in the order that they appear in the table.

Contact service personnel or Dräger when the remedy information indicates a service task, or if the symptom remains after all remedy actions have been
attempted.

5

Maintenance

5.1

Maintenance table

Service and test the breathing apparatus, including out-of-use apparatus, in accordance with the maintenance table. Record all service details and testing.
Refer also to the Instructions for Use for the lung demand valve, face mask and other associated equipment.

Additional inspection and testing may be required in the country of use to ensure compliance with national regulations.

Symptom

Fault

Remedy

High-pressure air leak or failed leak test

Loose or dirty connector

Disconnect, clean and reconnect couplings and 

retest

Faulty hose or component

Substitute user replaceable accessories and 

retest

Air leak from medium-pressure hose connection 

at the pressure reducer (safety relief valve)

Faulty O-ring, retainer, spring or pressure 

reducer

Service task

High or low medium pressure

Pressure reducer fault

Service task

Poor sounding whistle

Whistle dirty

Clean whistle flute and retest

Whistle not functioning correctly

Activation mechanism fault

Service task

Difficulty connecting or disconnecting the 

medium-pressure quick coupling

Dirty connector

Disconnect, clean and reconnect couplings and 

retest

Burring of the male coupling

Replace the hose with the male coupling

Component/System

Task

Before 

use

After 

use

Every 

month

Every 

year

Every 

10 

years

Complete apparatus

Visual inspection (see Note 1 and Section 3.5.1)

Notes

O Dräger recommendations

1

Clean the equipment if it is dirty. If it the equipment has been exposed to contaminants, disinfect any components that come into direct and prolonged
contact with the skin.

O

O

O

Functional testing (see Section 3.5.3)

O

O

O

Breathing cycle and static tests (see Note 2)

2

These maintenance tasks may only be carried out by Dräger or trained service personnel. Details of the tests are contained in the Technical Manual
which is issued to service personnel that have attended a relevant Dräger maintenance course.

O

Lung demand valve

Check push-in type connectors for lubricant (see Note 3)

3

For type A check the O-ring on the lung demand valve; and for type ESA check the outer surface of the male part of the push-in connector on the
lung demand valve. As a guide, lubricant should be felt on the fingers but not seen. If relubrication is required, lightly apply Dow Corning

®

Molykote

®

 111 (other lubricants are not tested and may damage the equipment).

O

Check the male element of the quick coupling for burring (see Step 3 in 

Section 3.1)

O

Pressure reducer

Medium-pressure check (see Note 2)

O

Inspect the sintered filter (see Note 2 and Note 4)

4

Replace the sintered filter if a drop in reducer performance is observed during a flow check or if it is visibly damaged.

O

Inspect the high-pressure connector O-ring (see Note 2 and Note 5)

5

Replace the high-pressure connector O-ring if it is found to leak during functional testing or if the O-ring is visibly damaged.

O

Overhaul. Contact Dräger for the Repair Exchange (REX) service 

(see Note 6)

6

Where the breathing apparatus is subjected to a high level of use (in training establishments etc.), reduce the overhaul period for the pressure reducer.
In these circumstances, Dräger recommend that the overhaul frequency should be less than 5,000 applications of use. An application of use is defined
as a single use of the fully assembled breathing apparatus, where the user breathes from the air cylinder. It does not include system pressurization
for pre-operational checks.

O

Cylinder

Charge cylinder to correct working pressure

O

O

Check charged pressure (stored cylinders only)

O

Check test date of cylinder

O

Recertification

According to national regulations in the country of 

use

Cylinder valve

Overhaul

At the time of cylinder recertification

!

Отзывы: