10.3 Changing the main electronics
1. Switch off the power to the device or declassify the area
according to the local regulations.
2. Loosen the set screw and unscrew the housing cover from
the device.
3. Lift the handle and remove the enclosure with the main
electronics.
4. Remove the protective caps for the connectors.
5. Unscrew the connector screws for the power supply and
relay and withdraw the connector.
6. Change the main electronics.
7. Plug the sensor cable and the connector for the power supply
and relay into the corresponding sockets and tighten them.
8. Re-fit the dust caps for the connectors.
9. Re-fit the sensor electronics and enclosure.
10. Screw the housing cover on again and tighten the set screw.
11. Switch on the power to the device.
12. Check and adjust the configuration as required.
13. Calibrate the device if necessary (see chapter 5 on page 15).
11
Disposal
Observe the applicable local waste disposal regulations.
For information consult your local environmental agency, local
government offices or appropriate waste disposal companies.
CAUTION
When the main electronics is changed, all settings and
parameters must be checked for correctness.
!
Disposal of electrical and electronic equipment:
EU-wide regulations governing the disposal of
electric and electronic appliances which have been
defined in the EU Directive 2002/96/EC and in
national laws have been effective from August 2005
and apply to this device. Common household
appliances can be disposed of using special
collecting and recycling facilities. However, as this
device has not been registered for household usage,
it must not be disposed of through these means.
The device can be returned to your national Dräger
Safety Sales Organization for disposal. Please do not
hesitate to contact the above if you have any further
questions on this issue.
Disposal of electrochemical sensors
In accordance with the material safety data sheet of
the corresponding sensor.
WARNING
Danger of explosions! Do not dispose of in fire,
risk of chemical burns! Do not open with force.
!