Reprocessing
72
Instructions for use
Polaris 600
Sterilization
For sterilization, use a vacuum steam sterilizer
(conforming to ISO 17665), preferably with
fractionated vacuum.
Only sterilize cleaned and disinfected items.
– Sterilize with a fractionated vacuum steam
procedure at 132 °C (270 °F) for 4 minutes.
– The sterilizable handle must not be subjected
to mechanical load during sterilization.
Otherwise, it may be permanently deformed.
– The sterilizable handle must not be reattached
until immediately before an operation.
Materials used in the light
The following materials are used in the light:
Assembling the parts
Fitting the sterilizable handle E
– see "Fitting the sterilizable handle E" on page
Fitting the sterilizable handle
– see "Fitting the sterilizable handle" on page 33.
WARNING
Risk of infection
If the sterilizable handle exhibits cracks or
deformation as a result of frequent
sterilization, the handle may fall into the
operation field and/or may not satisfy the
infection prevention policy.
Make sure that the handle is not used in the
event of such damage.
Components
Material
Light head with housing for light
control panel
Aluminum, pow-
der coating
Seals
Silicone
Bottom glass
Polycarbonate
(PC)
Sterilizable handle E
Polyetherimide
(PEI)
Sterilizable handle
Polyetherimide
(PEI)
Gimbal mounting
Steel, powder
coating
Integrated handle
Polyamide (PA)
Light control panel
Polyamide (PA),
polyester foil
Polaris 600 remote control for
MedView camera
Aluminum, pow-
der coating,
polyester foil
Components
Material