Assembly and preparation
32
Instructions for use
Polaris 600
of the remote control, the battery must be
replaced immediately, see "Replacing the
battery in the Polaris 600 remote control for
MedView camera" on page 58.
– Status LED does not light up: the battery is
completely discharged or is defective. To
reestablish the function of the remote control,
the battery must be replaced immediately, see
"Replacing the battery in the Polaris 600
remote control for MedView camera" on page
Fitting the handle
Fitting the sterilizable handle E
Immediately before an operation, fit the handle
under sterile conditions as follows:
1
Fit the handle (A) on the inner handle so that
the markers (B) line up.
2
Turn the handle (A) clockwise until the locking
mechanism (C) engages audibly.
3
Check that the handle is securely attached.
Turn the handle counterclockwise to do this.
26561
A
B
26
562
WARNING
Risk of patient injury
The sterilizable handle E can fall down into
the operating field if it is not engaged
properly.
The handle must be checked for secure
attachment after it has been fitted to the inner
handle.
C
A