20
Dräger X-am 5100
Use
3.1.5
Before entering the workplace
1. Switch on the instrument. The current measured values are shown
in the display.
Observe any warning
» «
or fault messages
» «
.
2. Check that the gas inlet opening on the instrument is not covered.
3.2
Configuration
To customise its configuration, the instrument must be connected to
a PC with the USB-DIRA adapter (Order No. 83 17 409).
The
Dräger
CC-Vision PC software is used to perform the
configuration.
Changing the configuration: see the Technical Handbook.
Standard instrument configuration
1
WARNING
Before any measurements relevant to safety are made, check
the adjustment with a bump test, adjust if necessary and
check all alarm elements. If national regulations apply,
a bump test must be performed according to the national
regulations. Faulty adjustment may result in incorrect
measuring results, with possible serious consequences.
The instrument can be operated normally. If the warning
message does not disappear automatically during operation,
the instrument must be serviced after the end of use.
The instrument is not ready to measure and requires
maintenance.
WARNING
Explosion hazard! To reduce the risk of flammable or
explosive atmospheres igniting, it is essential that the
warning notices below are observed:
In an oxygen enriched atmosphere (>21 Vol.-% O
2
), the
explosion protection cannot be guaranteed; remove
instrument from the Ex area.
High values outside the display area indicate an
explosive concentration where applicable.
!
!
NOTICE
Only trained personnel are permitted to make changes to the
device configuration.
Dräger X-am 5100
Bump test mode
Quick bump test
Fresh air adjustment
On
Life signs (visual only)
On
Disabling
permitted / blocked for A2
Averaging time
15 minutes for STEL
8 hours for TWA
1
Different settings can be selected to meet customer requirements on
delivery. The current setting can be checked and changed with the Dräger
CC Vision software.
i
i
Содержание MQG 0020
Страница 2: ......
Страница 28: ...28 Dr ger X am 5100 Technical Data...
Страница 69: ...Dr ger X am 5100 69 Dados t cnicos...
Страница 158: ...158 Dr ger X am 5100 1 6 170 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 163: ...Dr ger X am 5100 163 3 1 5 1 2 Bump Test 21 O2...
Страница 166: ...166 Dr ger X am 5100 166 3 6 3 6 1 A1 A1 A1 1 2 3 6 2 A2 A1 i i OK...
Страница 169: ...Dr ger X am 5100 169 5 2 1 68 06 291 2 3 4 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 1 Dr ger PTFE i i OK OK OK OK OK OK...
Страница 170: ...170 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Страница 220: ...220 Dr ger X am 5100 1 6 232 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 225: ...Dr ger X am 5100 225 3 1 5 1 2 Bump Test 21 O2...
Страница 227: ...Dr ger X am 5100 227 A1 A2 8 A1 230 3 4 A1 Dr ger X am 5100 1 0 5 l min 2 68 06 291 3 4 5 6 230 7 8 A1 OK OK...
Страница 228: ...228 Dr ger X am 5100 9 A1 3 5 228 3 6 3 6 1 A1 A1 A1 A1 A2 3 6 2 A2 A1 i i OK...
Страница 230: ...230 Dr ger X am 5100 4 2 4 3 1 Dr ger CC Vision 2 60 5 5 1 0 1 2 3 3 4 OK 5 5 Bump Test Bump Test 226 230 OK OK OK OK...
Страница 231: ...Dr ger X am 5100 231 5 2 1 68 06 291 2 3 4 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 1 m Dr ger PTFE i i OK OK OK OK OK OK...
Страница 232: ...232 Dr ger X am 5100 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr ger 6 2 7 Dr ger Dr ger...
Страница 259: ...Dr ger X am 5100 259 1 6 271 Dr ger Dr ger Dr ger Dr ger Dr ger Dr ger 1 1 i i...
Страница 264: ...264 Dr ger X am 5100 3 1 4 3 2 1 3 1 5 1 2 Bump Test Bump Test OK 21 Vol O2...
Страница 270: ...270 Dr ger X am 5100 5 5 6 001 7 8 9 10 11 12 OK 13 5 14 OK OK OK OK OK OK...
Страница 271: ...Dr ger X am 5100 271 6 6 1 EN 60079 29 2 EN 45544 4 4 Dr gerSensor 6 2...
Страница 286: ...286 Dr ger X am 5100 Notes on Approval A 2 Declaration of Conformity...
Страница 287: ......