Instructions for Use Tropic
plus
Symbols
Intended use
Tropic
plus
– ultrasonic nebulizer
– Used to humidify breathing air for patients who are breathing
spontaneously. The humidified air is free-flowing and directed
towards the patient.
– For aerosol therapy, sputum-dissolving substances can be
added to the nebulizer.
Environment of use
The medical device must is only to be used in clinical rooms.
Assembling accessories
Strictly observe assembly instructions and instructions for use.
Table-mounted version
Fit the pole (300 mm long) into the holder at the back of the
nebulizer and secure it in place with the locking screw.
Slide the holding-down device (to secure the lid of the nebu-
lizer chamber) and the hinged arm onto the pole and position
them as required.
Caution
Warning! Strictly follow
these instructions for
use
Disposal informa-
tion
Manufacturer
Atmospheric pres-
sure
Date of manufacture
Relative humidity
Storage temperature
limitation
NOTE
Aerosol therapy must only be carried out as directed by a med-
ical professional.
WARNING
Risk of explosion
This medical device is neither approved nor certified for
use in areas where the oxygen concentration may exceed
25 Vol % or combustible or explosive gas mixtures are like-
ly to occur.
WARNING
Risk of explosion
Do not use the device together with combustible or explo-
sive liquids.
WARNING
Risk of patient injury
Do not use the ultrasonic nebulizer on patients requiring
ventilation.
WARNING
Risk of patient injury
Do not install the ultrasonic nebulizer in breathing circuits.
WARNING
Risk of patient injury
Do not use the ultrasonic nebulizer with a ventilation mask,
tracheal cannula, or mouthpiece.
This can lead to accumulations of liquid in the respiratory
system.
WARNING
Risk of infection
Do not use the ultrasonic nebulizer without a bacterial filter.
NOTE
The air flow is only filtered adequately if a suitable bacterial filter
is used.
Replace the bacterial filter after 24 hours of operation or in
accordance with the special instructions provided by the respec-
tive filter manufacturer.
Reusable hoses must be cleaned once a day or whenever there
is a change of patient.
Disposable hoses must be replaced once a day or whenever
there is a change of patient.
CAUTION
Installation to the basic device must be in accordance with the
instructions for use of the basic device on which this medical
device is used. Make sure that the connection to the basic
device system is secure.
002