Pocket Guide
Evita XL
9
Preparation
Preparation
How to start the
Evita XL
Preparing the
Evita XL
– Connect the power supply
– Connect the gas supply (Air + O
2
)
– Install expiratory valve, flow sensor
– Connect the breathing circuit
– Switch on
Evita XL
– Activate Standby mode
– Enter the humidification type
Checking readiness for operation
– Perform the Device Check
– Perform the Airtight Check
– Test the DC power pack (batteries)
Selecting patient and therapy settings
– Select the patient
– Set the ventilation parameters
– Set the alarm limits
Starting ventilation
– Touch start button in the Start/Standby dialog win-
dow
Connecting to the power supply
The mains voltage must correspond to the voltage ran-
ge indicated on the rating plate.
z
Insert plug into the mains power socket.
For operation with DC power pack and external battery
(DC power pack option):
z
Connect optional external battery with cable.
Connecting to the gas supply
1
Screw on the compressed air hose (A) to the Air inlet
connector and the compressed O
2
gas hose (B) to
the O
2
inlet connector on the back panel of
Evita XL
.
2
Connect the plugs to the central gas supply wall ter-
minal units.
NOTE
The
Evita XL
Pocket Guide is not a replacement or substitute for the Instructions for Use.
Any use of the device requires full understanding and strict observation of the Instructions for Use.
Either:
220 V to 240 V
or:
100 V to 127 V
006
Air O
2
A