Technical data
Technical data
Ambient conditions
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
During operation
Altitude
Up to 3000 m (9842 ft)
Set values
Applies to:
– Evita Infinity V500
– Babylog VN500
– Evita V300
The desired parameters can be set without loss of
accuracy using the therapy controls. The control-
led parameters – pressure, flow, volume, and O
2
concentration – can only be applied with the accu-
racy of the associated measured values.
The accuracies indicated apply only under the fol-
lowing conditions:
– The device is ready for operation, see chapter
"Getting started."
– Any accessories being used are approved for
the device, see the list of accessories.
– The type of humidification is selected correctly
in the
Start/Standby
>
Br. circuit/ Humidifier
dialog window.
The tolerances do not include the measurement
uncertainty of external test equipment. This infor-
mation is available on request.
O
2
concentration
T0 ... 90
Test conditions as per ISO 80601-2-12:2011,
Sec. 201.12.1.104
Adults
(Evita Infinity V500, Evita V300)
<35 s