Alarm – Cause – Remedy
24
!
201
Battery activated
The ventilation unit is
powered by the battery as
there is no mains power
supply.
– If PS500 is not con-
nected, the maximum
operating time is
30 minutes.
– If PS500 is connected,
the maximum operating
time is 360 minutes.
Connect device to the
mains power supply.
!
127
Battery charging deferred
Battery charging is deferred
to prevent battery
overheating. The device
can be used normally.
Battery charging continues
automatically and is
indicated by a flashing
segment in the battery
symbol.
!!!
254
Battery discharged
The remaining operating
time of the battery is less
than 5 minutes.
Connect device
immediately to the mains
power supply.
!!!
250
Battery low
The battery is almost
discharged.
Connect device to the
mains power supply.
!!
251
Battery low
The battery is almost
discharged.
Connect device to the
mains power supply.
!
100
Continuous nebulization
activated
Continuous nebulization
was activated by the user.
To end continuous
nebulization, press the
"Cancel" button if required.
!!!
253
Device failure
A system failure was
detected.
Disconnect patient from the
device and continue
ventilation without delay
using another independent
ventilator.
Call DrägerService.
!!!
253
Device failure (10)
A failure was detected by
the safety software system.
Disconnect patient from the
device and continue
ventilation without delay
using another independent
ventilator.
Call DrägerService.