14. Maintenance and Service
14.1 Storing
El.-chem. gas sensors react to gases in ambient conditions, even if the instrument is
switched off. Make sure, that the instrument is stored in a place with room temperature with-
out contamination with solvents, exhaust gases or combustibles and that it becomes re-
charged periodically.
14.2 Maintenance
The gas conditioning cartridge should be cleaned after been in use. In addition to that, the
cartridge should be checked due to tightness (O-ring seal). The filter discs and fleece have
to be changed if dirty (see 3.4 too).
The housing of the instrument can be cleaned with a damp cloth. Take care that the gas
outlet at the bottom of the instrument will not get blocked.
14.3 Service
In order to assure accurate measurement and the reliability of the functions the MSI EM200
E should be checked according to requirement of EN 50379 and if applicable be calibrated
by an authorised service point once a year.
14.4 Consumables and accessories
Printer with infra-red data transmission 5600401
Paper for IR-Printer
5690151
Consumable set 2
5600411
10 x disc filter
20 x filter fleece ø 26 mm
Plug-in charger P7/EM200
5680106
Wall power supply (Draeger 8315805) 5680111