Troubleshooting
Dräger DrugTest 5000 Test System
23
22001
(QA
Check
failed)
QA Check failed
Device cannot be used. Maintenance is
required.
QA-Test failed! Remove cassette.
Close door.
20006
Test cassette cannot be
processed
Ensure correct QA cassette is used
Error when reading QA-cassette!
Remove cassette.
22020
Login as operator failed
(wrong password was
entered)
Enter correct password
Invalid password. Login cancelled.
22040
New operator: operator list full Delete one operator
Too many operators.
No further entry possible.
Transfer new operator list.
22041
New operator: operator
already exists
Use another name
The operator has already been set up
and could not be stored again.
22043
New operator: password
already used or 2nd pass-
word is 1st password
Choose a different password or enter
correct new password
Invalid password.
Please choose another password.
22050
Change operator password:
wrong old password
Enter correct password
Wrong password entered.
Please try again.
22052
Change operator password:
new password already used
Choose different password from existing
one
Invalid password.
Please choose another password.
22053
Change operator password:
2nd new password is 1st new
password
Enter correct new password
Two different passwords have been
entered.
Enter password again.
New operator not assigned.
30000
Service: Service expired
Device can be used, but maintenance
should be carried out soon.
"Service 1" expired!
Please carry out maintenance.
30001
Service: Service to be carried
out soon
Device can be used, but maintenance
should be carried out soon.
"Service 1" to be carried out soon.
30013
User error: Measurement
cancelled by operator
No error
Do you really want to cancel the current
measurement?
Cancel measurement by pressing OK or
return to measurement.
30022
User error: Temperature
inside of the instrument is out
of working range
Usage within temperature range
according to IFU
Device temperature is beyond operational
range.
Wait for equalization of temperature or
switch off device.
Error
Code
Description
Corrective Action
Text on Screen