Basics
Dräger X-dock 5300/6300/6600
19
2. Select desired gas inlet.
The configuration menu will appear.
When using a Dräger gas cylinder:
1. Enter the Part No. of the Dräger gas cylinder.
All the necessary details for the configuration will be
automatically filled in. The batch number and the expiry
date can be entered manually in addition.
2. If necessary select
Further options
and , to reset the
gas cylinder level indicator.
3. If required, configure other gas inlets in the same way.
When using a gas cylinder from another manufacturer:
1. Create or delete test gas component.
Create a new test gas component with .
Delete the current test gas component with .
2. Select gas.
3. Enter the gas concentration.
4. Select gas units.
5. If required, create additional test gas components.
6. The following information can optionally be given:
Part No. of the gas cylinder
Lot number of test gas cylinder
Use-by date of the gas cylinder
7. If required, enter
Further options
.
8. For more information see the X-dock 5300/6300/6600
Technical Manual.
4.6.1
Settings
1. Select
>
Station gas configuration > Settings
.
The OV sensor of the PAC 7000 OV allows to choose between
two different test gases that are also used for calibration and
testing. These test gases are carbon monoxide (CO) and
ethylene oxide (EO).
For the X-am 2000, it is possible to choose between 3 different
test gases that are also used for calibration and testing. The
three options are methane (CH
4
), propane (C
3
H
8
) and
pentane (C
5
H
12
). The sensor is calibrated with different
sensitivity levels depending on the gas selected. More
information on this can be found in the relevant sensor data
sheets.
Select the gas inlet
►
Select Test gas inlet
►
Enter part no.
►
◄
Enter batch no.
Enter Expiry date
►
◄
Create Test gas
component
Select Test gas
►
◄
Delete configuration
Enter concentration
►
◄
Select unit
Further options
►
NOTICE
Upon entering the part no. of a Dräger gas cylinder, a
gas cylinder level indicator is displayed automatically,
unless this function was deactivated before (see
NOTICE
The values entered automatically must be matched
with those indicated on the test gas cylinder. If these
values are not identical, the value indicated on the gas
cylinder is to be considered and a manual correction of
the values must be made.
i
i
i
i
NOTICE
Deleting all test gas components deletes all
information of the test gas inlet.
Select test gas for X-am 2000
►
Fresh air inlet setting
►
Expert settings
►
Select test gas for PAC 7000 OV
►
Gas cylinder level indicator
►
NOTICE
The corresponding gas must be connected to one of
the gas inlets and set in the gas configuration.
i
i
i
i