Dräger Bodyguard 7000 Скачать руководство пользователя страница 2

Bodyguard

®

 7000

Electronic monitoring system

,QVWUXFWLRQVIRUXVH

$'33DJH

3.7  Product marking and symbols

3.7.1 Serial number

The product serial number is marked on the pressure module and user 
interface in the format BRMY-1234 MM/YY, where MM/YY are the month 
and year of manufacture. The pressure module has an additional 5-digit 
suffix which is the serial number of the internal HUD transmitter unit.

3.7.2  LCD screen symbols

4 Operation

4.1  Preparation for use

NOTICE

The LCD screen has an external sacrificial screen which is a removable 
clear screen. The sacrificial screen might be fitted with a thin flexible 
protective covering. Remove this thin covering before first use. Do not 
remove the sacrificial screen.

1. Carry out a visual inspection (see section 6.2).
2. Assemble and prepare the breathing apparatus for use as described 

in the product instructions for use.
– When instructed to carry out functional testing of the breathing 

apparatus, carry out the Bodyguard

®

 7000 functional testing (see 

section 6.3).

3. Put on the breathing apparatus as described in the product 

instructions for use.
– When the cylinder valve is opened, the self-test and start-up 

sequences run and then the system adopts the active mode.

4. DSU automatic distress alarm:

– Tally version: remove the tally to activate the motion sensor.
– Button version: remove the tally if the motion sensor is required.

4.2 During 

use

4.2.1 User functions

 All electronic devices could suffer a temporary loss of function if 

subjected to high levels of RF radiation. The system operates with no 
loss of performance or function once the RF radiation has been removed.

– Regularly check the LCD screen to confirm the cylinder pressure and 

TTW.
– Cylinder pressure and TTW are indicated digitally on the normal 

operating screen (Fig. 3). Approximate cylinder pressure is also 
indicated by the radial segments on the screen.

Fig. 3

Normal operating screen

– To call for emergency help or assistance, press the manual alarm 

button .

– To illuminate the screen backlight, press and quickly release the left-

hand or right-hand button.
– Pressing the right-hand button also scrolls any stored user ID 

information.

4.2.2  Alarms and status messages

 WARNING

Users should be in a safe area before the EOST alarm and mechanical 
whistle commence.

Evacuate to a safe area immediately if warnings commence during an 
operation.

– The green LED on the user interface flashes every second to indicate 

active mode.

– If the motion sensor pre-alarm activates, and is not required, move 

the user interface to cancel the alarm. Do not use the buttons to 
switch off the pre-alarm.

– If the motion sensor full alarm activates, cancel the alarm as follows:

– Tally version: fit the tally. With the tally fitted, the system is not in 

active mode. It is in a non-operational mode and there is no 
display of time to whistle.

– Button version: simultaneously press and hold the left-hand and 

right-hand buttons until the alarm stops.

– If a low battery   or battery error   displays, an alarm tone sounds 

every 9 seconds.
– When a low battery is first indicated, it is possible to safely use the 

breathing apparatus for up to 2 hours. However, Dräger strongly 
recommend replacing the batteries or recharging the power pack 
at the first opportunity (see section 6.5).

– If a battery error displays, evacuate to a safe area immediately.

4.2.3 Alarm patterns

Pre-alarm pattern

An increasing-volume 3-tone alarm is emitted from the user interface, 
and the LCD screen backlight flashes.

Full alarm pattern

A high-pitched repeating alarm tone is emitted from the user interface, 
and the red and blue LEDs on the user interface flash.
– Symbol 

 displays during automatic distress alarms.

– Symbol   displays during manual distress alarms.

EOST alarm pattern

An intermittent high-pitched alarm is emitted from the user interface. The 
red and blue LEDs on the user interface flash, and a sector on the left of 
the LCD screen flashes red. The mechanical whistle on the breathing 
apparatus also commences at approximately the same time.

4.2.4  Telemetry and retreat alert

The symbols used during telemetry (PSS

®

 Merlin

®

) and retreat alert 

operations are shown in section 3.7.2.
– See section 4.5.5 for a description of retreat alert operation and 

symbols.

– See the PSS

®

 Merlin

®

 instructions for use for a description of 

telemetry operation and symbols.

4.3 After 

use

 WARNING

Removing the breathing apparatus in a hazardous breathing environment 
is unsafe.

Do not remove the breathing apparatus until in a safe breathing 
environment.

1. Remove the breathing apparatus as described in the instructions for 

use supplied with the breathing apparatus.

2. Switch off the electronic system (see section 4.4.5).
3. Carry out the after use tasks detailed in the maintenance table (see 

section 6.1).

4.4  Basic operating functions

4.4.1  Self-test and start-up sequences

The system switches on when the power pack is fitted, when the left-
hand button is pressed, when the tally is removed, or when the 
pneumatic system is pressurized. Each time it switches on (with or 
without an air cylinder fitted), the self-test and start-up sequences run.

4.4.2 Self-test

During the self-test, alarm tones sound, the LEDs and screen flash, and 
the start-up sequence runs.
– If the system passes the self-test, it adopts the active mode (see 

section 4.4.4).

– If the system fails the self-test, one of the following faults is indicated.

– The screen displays a cross   and a fault code. Note the fault 

code and contact Dräger.

– The screen displays a low battery   or battery error   and 

switches off. See troubleshooting (section 5) for remedy actions.

4.4.3 Start-up sequence

The start-up sequence provides information and options for the user each 
time the system switches on. The sequence is configurable using the 
Dräger PC Link (see section 3.3.1), and only the pre-configured options 
and information are shown during start-up.

Default start-up sequence

The start-up sequence shown in this section is the default setting for 
Bodyguard

®

 7000.

1. The wait symbol   and then pass symbol   displays as the system 

performs the self-test.

2. The cylinder type symbol   or   displays and the selected cylinder 

type is shown. If more than one cylinder type is stored in the system 
memory, the user can select the required cylinder from the available 
options (see section 4.5.3).

3. Electronic leak test   displays to allow the user to check the 

breathing apparatus pneumatic system for leaks (see section 6.3.1).

4. Motion sensor activated 

 or deactivated   displays to indicate the 

status of the motion sensor of the automatic distress alarm.

Other configurable start-up options

– User ID   allows information about the wearer to be uploaded from a 

user ID card to the system memory (see section 4.5.2).

– Battery condition   indicates the charge state of the power pack 

using the radial segments on the user interface screen. Battery 
condition is also referred to as the “fuel gauge”.

– Quiet alarm   allows the user to select reduced volume alarms for 

operating inside a restricted space such as a chemical protective suit 
or a repair workshop (see section 4.5.4).

4.4.4 Active mode

Active mode is the main operating mode of the electronic monitoring 
system. When the system is in active mode, the electronic monitoring 
and warning functions are operational. Active mode is indicated by the 
following.
– The normal operating screen (Fig. 3) displays on the user interface.
– The green LED on the user interface flashes every second.

 If the tally is still fitted in a tally version, the system is not in active 

mode. It is in a non-operational mode: TTW is not displayed on the LCD 
screen, but cylinder pressure is displayed.

4.4.5  Switching off the electronic system

The system cannot be switched off unless the cylinder pressure indicated 
on the user interface is below a preset value. Close the cylinder valve 
and fully vent the pneumatic system before attempting to switch off.

Tally version

– After normal use with the tally removed: refit the tally.
– If the system was activated with the tally fitted (during functional 

testing for example): press and hold the left-hand button until the 
return symbol   momentarily displays, then immediately release the 
button.

Button version

– Press and hold the right-hand and left-hand buttons until the display 

clears, then immediately release the buttons. If the tally was removed, 
refit the tally.

4.5  Additional start-up and operating functions

4.5.1 General

The functions in this section are only available if they are pre-configured 
in the system. More information about configurable settings and features 
is in section 3.3.1.

4.5.2  User ID (scrolling data)

Information about the wearer (e.g. user name, brigade name, and station 
number) can be uploaded from a user ID card to the system memory. 
Once information is stored, pressing the right-hand button during use 
scrolls the information across the screen. The scroll speed is configurable 
using the Dräger PC Link (see section 3.3.1).

Uploading information from a user ID card

1. Hold the user ID card on to the back of the user interface, directly 

behind the screen.

2. Press the left-hand button to activate the start-up sequence.
3. When the user ID symbol   displays, press the left-hand button 

before all of the radial segments extinguish.
– The screen briefly illuminates, and the system uploads data from 

the user ID card to the system memory.

– When the data upload is complete, the tick symbol   displays.

4. Press and release the right-hand button to check that the uploaded 

data is correct.
– The user ID scrolls from right to left across the screen.

5. Repeat the procedure if the information is incomplete or inaccurate.
6. Switch off the electronic system (see section 4.4.5) if necessary.

4.5.3 Cylinder selection

If more than one cylinder type is stored in the system memory, the user 
can select the required cylinder type during start-up as follows.
1. Press the left-hand button to activate the start-up sequence.
2. When the cylinder type symbol   or   displays, press the left-hand 

button before the radial segments extinguish.
– The screen displays the next stored cylinder type and the radial 

segments extinguish clockwise.

– The cylinder size and pressure alternate on screen (Fig. 4 shows 

a 9.0 litre, 300 bar cylinder).

Fig. 4

Cylinder selection screen

3. Before the radial segments extinguish, press the right-hand button to 

scroll between the stored cylinder types.

4. When the required cylinder type displays, immediately press the left-

hand button to confirm selection.
– The start-up sequence restarts, and the selected cylinder type is 

shown during the sequence.

5. Switch off the electronic system (see section 4.4.5) if necessary.

4.5.4 Quiet alarm

Quiet alarm allows the user to select reduced volume alarms for 
operating inside a restricted space such as a chemical protective suit or a 
repair workshop. Select quiet alarm as follows.
1. Press the left-hand button to activate the start-up sequence.
2. When the quiet alarm symbol   displays, press the right-hand button 

before the radial segments extinguish.
– The alarm volume is reduced for the current operation only. The 

system automatically resets to full alarm volume when it is 
switched off.

General symbols

Low battery

Press right-hand button

Battery or charging error

Wait

Battery charged

Leak test timing

Low cylinder pressure

Leak test pass / vent system

Manual distress alarm

Leak test fail

Automatic distress alarm

User ID

Passed or complete

Cylinder type – Single

Failed or cancelled

Cylinder type – Double

Perform leak test

Reset or return

Open valve

Quiet alarm

Close valve

PC Link mode

Telemetry symbols

Active communication

Fault

Lost communication

Evacuate

Voluntary withdrawal

Retreat alert symbols

TTR

Arrival point

Retreat

41221

bar

41253

bar

Отзывы: