Prepare the cleaning agent and the 

disinfectant in accordance with the 

manufacturer’s instructions.

Place the components in the 

cleaning agent.

Ensure that the solution flushes 

all surfaces and interior spaces.

Swirl the components back and 

forth at least 3 times. Make sure 

the cleaning agent reaches all 

surfaces and interior spaces.

Fill a syringe that can hold at least 50 mL with the cleaning agent. Remove the

flow sensor housing from the cleaning agent and spray at least 50 mL through

each opening. Put the flow sensor housing back into the cleaning agent.

*  The Neonatal Flow Sensor can also be used with Evita

®

 V-Series.

Reprocessing of ISO 15 Neonatal Flow Sensor  
and Y-piece Neonatal Flow Sensor*

D

-5755-2018

Dräger Babylog® V-Series  
Dräger Neonatal Flow Sensor 

Reprocessing instructions 
Manual cleaning

Sensor housing of the flow sensor Y-piece

Sensor housing of the ISO 15 flow sensor

At the end of the specified contact time, swirl the components 

back and forth again at least 3 times.

Sensor housing of the flow sensor Y-piece

Sensor housing of the ISO 15 flow sensor

This guideline does not replace or amend the instructions for use. Please always refer to the 

reprocessing instruction in the latest instruction for use of the components. The instructions 

for use and the hospital guidelines and policies always have to be observed.