Apollo
Safety Features
10.3 Gas Supply Failure
If the gas supply fails, the Apollo displays a message in the status field at the top of the screen,
corresponding to the failed gas:
“NO AIR SUPPLY”
“NO N
2
O SUPPLY”
or “NO O
2
SUPPLY”
Open the valve on the corresponding backup cylinder at the back of the machine
Restore the central pipeline gas supply
If there is no cylinder backup supply for the failed gas, the corresponding LED on the front panel
of the machine will flash red.
In case of N2O or air failure, 100% O2 should be delivered; be sure to set the O2 to an
appropriate flow.
In case of O2 failure, the SORC prevents hypoxic gas mixtures.
Oxygen must be restored immediately.