
|
Apollo
Safety Features
10.9 System Failure
If the system no longer responds to an action:
1. Ventilate the patient manually.
2. Switch the machine off and on again.
3. Cancel the self test.
4. Check the vaporizer setting.
5. Close all flow controls (except O2), and run 100% O2.
If the breathing bag does not fill with fresh gas or manual ventilation is not
possible:
• check the oxygen supply; open cylinder valves if necessary
If fresh gas is still not delivered:
• close all flow controls
• disconnect patient from workstation and use another device
• turn workstation off