Safety-related information

6

en

2

Safety-related information

Strictly follow the instructions for use.

Any use of the instrument and its accessories requires full 
understanding and strict observation of these instructions.

The instrument is only to be used for the purposes specified 
herein.

We recommend that a service contract be obtained with Dräger 
Safety and that all repairs be carried out by them. 

Liability for proper function or damage

The liability for the proper function of the instrument is irrevocably 
transferred to the owner or operator to the extent that the instrument is 
improperly serviced or repaired by personnel not employed or 
authorized by Dräger Safety or if the instrument is used in a manner 
not conforming to its intended use.
Dräger cannot be held responsible for damage caused by non-
compliance with the recommendations given above.
The warranty and liability provisions of the terms of sale and delivery 
of Dräger are likewise not modified by the recommendations given 
above.
Dräger Safety AG & Co. KGaA

Содержание Alcotest 6820 GB / SCT

Страница 1: ...i Alcotest 6820 GB SCT Breath Alcohol Screening Device with Datamanagement Instructions for Use ...

Страница 2: ...vals 12 4 4 Explanation symbols 12 5 Use 13 5 1 Intended use 13 5 2 First usage 13 5 3 Prerequisites 13 5 4 Preparation for use 14 5 5 Performing the measurement 14 5 6 Results 15 5 7 Further measurement 16 5 8 Refusing a test 16 5 9 After sampling 17 6 Set up the device 18 6 1 Switching the device on and off 18 6 2 Straps 18 6 3 Batteries 19 6 4 Mouthpieces 20 7 Menu 20 7 1 Menu symbols 21 7 2 Na...

Страница 3: ...u or menu 21 8 Modes 21 8 1 Mode settings 22 8 2 Changing the mode 22 9 Troubleshooting 23 10 Maintenance 25 10 1 Cleaning 25 10 2 Accuracy check 25 10 3 Calibration 26 10 4 Storage 26 10 5 Disposal 27 11 Technical data 28 12 Order list 31 ...

Страница 4: ...trations and Screens 4 en 1 Illustrations and Screens 1 1 Illustrations The Illustrations help to visualise the tasks described in some sections 00233765 Hand strap Wrist strap Attaching mouthpiece Batteries ...

Страница 5: ...LERT FAIL 100 μg 100 ml ZERO PASS 0 μg 100 ml SKIP Subject Ethnic SKIP Subject Ethnic W1 White British Menu Back to menu Check value 35μg 100 ml Gas Type Dry gas Accuracy check Next Back to measurement Last test Set contrast Last cal date Set mode Accuracy check Providing Refusal REFUSAL YES NO Invalid results Insufficent Volume OK SKIP Subject Ethnic W1 OK SKIP Subject Ethnic W1 OK ...

Страница 6: ...or proper function or damage The liability for the proper function of the instrument is irrevocably transferred to the owner or operator to the extent that the instrument is improperly serviced or repaired by personnel not employed or authorized by Dräger Safety or if the instrument is used in a manner not conforming to its intended use Dräger cannot be held responsible for damage caused by non co...

Страница 7: ...at require a greater awareness by the user A definition of the meaning of each icon is as follows CAUTION Indicates a potentially hazardous situation which if not avoided could result in physical injury It may also be used to alert against unsafe practices NOTICE Indicates a potentially hazardous situation which if not avoided could result in damage to the product or environment ...

Страница 8: ...Description 8 en 4 Description 4 1 What s in the box 1 Dräger Alcotest 6820 measuring device incl 2 batteries 2 Mouthpieces 3 pieces 3 Wrist strap 1 3 2 00833765 ...

Страница 9: ...Description 9 en 4 LED red green yellow 5 DOWN MENU button 6 Charging communications socket 7 OK button 8 UP button 9 Display 10 Screw plug 00733765 5 8 4 7 9 10 6 ...

Страница 10: ... function symbols For menu related symbols on page 19 section Menu symbols Button Function Switching on and off the device and confirming entries or selected functions In further text written OK Entering the menu and navigating down in the menu and jump between digit positions In further text written DOWN or MENU Navigating up in the menu jump between digit positions In further text written UP Sym...

Страница 11: ...ours e g disinfectants 5 3 1 Adequate breath sample The test person should breath regularly and calmly before the sample is taken Avoid breathing in and out quickly through the mouth The test person must be capable of exhaling the minimum breath volume For a valid test the exhaling breath must constantly flow for at least 2 seconds factory settings Before sampling the test person must breathe norm...

Страница 12: ...he required input If the device is ready for measurement READY is shown in the display 6 Insert a new mouth piece on page 18 section Mouthpieces An new mouth piece must be used for each test 5 5 Performing the measurement 1 Make sure all prerequisites on page 11 section Prerequisites are met and all preparations on page 12 section Preparation for use are done 2 Ask the test subject to blow evenly ...

Страница 13: ...resentation The test person has got an unobtrusive alcohol level It can be displayed the following massages ALERT FAIL or the value in a decimal presentation The test person has got an noticeable alcohol level Continue aton page 15 section After sampling A sampling with low volume can cause an error In this case the horn sounds briefly the red LED flashes and the display shows Insufficient Volume ...

Страница 14: ...covery time the display shows WAIT 3 Continue at 5 5 Performing the measurement 5 8 Refusing a test If the test person is unable or unwilling 5 8 1 Providing REFUSAL 1 press UP to complete the test The result is detected as REFUSAL 2 To proceed the result confirm with YES To cancel the result press NO 5 8 2 To repeat the REFUSAL 1 press OK The instrument is ready in 4 seconds ...

Страница 15: ...ntity 1 Enter by pressing OK 2 Set digits by using the UP and DOWN buttons To enter a 5 digit code use Blank for the first digit 3 Confirm by pressing OK 4 Confirm by pressing OK twice Location of Test 1 Enter by pressing OK 2 Set digits by using the UP and DOWN buttons 3 Confirm by pressing OK 4 Confirm by pressing OK twice If the input is done successfully Data saved is shown in the display and ...

Страница 16: ...on Performing the measurement Switch off the device Switch off during Test Reason status 1 Enter by pressing OK 2 Set Subject Age and Gender READY is displayed Now it s possible to switch off the device by pressing OK for approx 4 sec The device will switch off automatically after 4 minutes without using 6 2 Straps To have the right grip on the device two different straps can be mounted See on pag...

Страница 17: ... with the screws 2x 5 Insert the bottom lug of the hand strap in the bottom opening on the device and fix it with the screws 2x of the hand strap 6 Close the battery lid and tighten the screw plug 6 3 Inserting or replacing the batteries Inserting the batteries upon initial operation or changing charging if the battery symbol shown in the displays right upper corner indicates empty See also on pag...

Страница 18: ...ached to the device With mouthpiece receptacle With the fluted markings press on surface facing upwards attach the mouthpiece at an angle push it upwards and press it into the mouthpiece receptacle The mouthpiece audibly snaps into place The mouthpiece can be used facing to the right or to the left 3 Remove all packaging The mouth piece is now ready for measurement 7 Menu The menu shows following ...

Страница 19: ...K 7 3 Leaving the sub menu or menu 1 Navigate to the upper entry by pressing UP several times The upper entry can be Back to measurement or Back to menu 2 Choose the entry by pressing OK 8 Modes The device is set up with 4 different modes listed in the menu on page 18 section Menu The menu shows following entries AIR ROAD MARINE RAIL Symbol Meaning Return to next higher menu level Back to measurem...

Страница 20: ...pressing OK AIR ROAD MARINE RAIL ZERO 4 μg 100 mL 4 μg 100 mL 4 μg 100 mL 4 μg 100 mL PASS 9 μg 100 mL 31 μg 100 mL 31μg 100 mL 31 μg 100 mL ALERT 35 μg 100 mL 35 μg 100 mL 35 μg 100 mL FAIL 9 μg 100 mL 36 μg 100 mL 36 μg 100 mL 36 μg 100 mL AIR ROAD MARINE RAIL ZERO 4 μg 100 mL 4 μg 100 mL 4 μg 100 mL 4 μg 100 mL PASS 9 μg 100 mL 17 μg 100 mL 20 μg 100 mL 31 μg 100 mL ALERT 23 μg 100 mL 26 μg 100...

Страница 21: ...o measurement mode after 120 sec without use Enter the menu again on page 18 section Menu can not be switched on empty batteries on page 17 section Inserting or replacing the batteries Displayed fault Cause Remedy Insufficient volume Invalid results on page 13 section Results on page 11 section Adequate breath sample Calibration error Calibration not successful on page 24 section Calibration Accur...

Страница 22: ... Accuracy check Service required accuracy check Accuracy interval expired Service date expired accuracy check service date expired Calibration overdue lockout in x days Calibration interval expired Perform calibration by trained service personal Service required Calibration Calibration period expired Service date expired calibration Service date expired Memory for XXTest Device can be used for XX ...

Страница 23: ...n on the display An accuracy check must be performed The accuracy check can be performed at any time Before performing the accuracy check the data must be saved on a PC Downloading the data of the device and clearing it s memory can be performed with Dräger Diagnostics software on page 28 section Order list software The accuracy check doesn t affect the calibration settings 1 Enter the menu on pag...

Страница 24: ...rror will be displayed if a download and memory clearing has not been completed 10 3 Calibration Alcotest 6820 GB is supplied with standard calibration by the manufacturer The date of the last calibration can be found in the quick menu 10 4 Storage The Alcotest 6820 GB consumes a small amount of current even when switched off CAUTION Leakage of battery acid can cause damage to the device Remove ba...

Страница 25: ...to open This product must not be disposed of as household waste This is indicated by the adjacent symbol You can return this product to Dräger free of charge For information please contact the national marketing organisations or Dräger Batteries must not be disposed of as household waste This is indicated by the adjacent symbol Dispose of batteries at battery collection centres as specified by the...

Страница 26: ...ve data a conversion factor of 2100 is used for breath and blood alcohol concentration data Units of measure g L mg 100 mL mg mg mL mg L μg 100 mL μg μg L g 210 L Ambient conditions During operation 0 C to 50 C 32 F to 122 F 10 to 100 rel hum non condensing 600 to 1300hPa During storage 20 C to 70 C 4 to 158 F Taking Samples Respiratory volume 1 2 L default Minimum flow rate 8 L min default Minimu...

Страница 27: ...alue whichever value is higher Sensitivity drift Typically 0 4 of the measured value per month Wait time for measurement readiness following a previous measurement with 25 μg 100mL 20 s 0 5 BAC 20 s 50 μg mL 40 s 1 0 BAC 40 s 100 μg 100mL 80 s 2 0 BAC 80 s 150 μg 100mL 120 s 3 0 BAC 120 s Dimensions 65 mm x 147 mm x 39 mm WxHxD Weight approx 260 g Power supply 2x1 5 V Type Alkaline Mignon LR6 AA B...

Страница 28: ...test 6820 GB 83 27 288 Alcotest 6820 SCT 37 03 695 Accessories Mouthpieces hygienically tested 100 pieces 68 10 690 250 pieces 68 10 825 1000 pieces 68 10 830 Alkaline battery 1 piece 13 35 804 Case for Alcotest 6820 GB 83 17 920 Leather pouch for Alcotest 6820 GB belt fixation 83 17 911 Power supply 83 16 991 Cable for PC connection RS232 83 19 715 Draeger Diagnostics software 83 17 915 ...

Страница 29: ...Order list 29 en ...

Страница 30: ......

Страница 31: ......

Страница 32: ...ax 49 451 882 20 80 www draeger com 90 33 765 IFU 4754 580_en Dräger Safety AG Co KGaA Edition 03 August 2021 Edition 01 August 2016 Subject to alteration Importer UK Draeger Safety UK Ltd Ullswater Close Blyth Riverside Buisness Park Blyth Northumberland NE24 4RG United Kingdom ...