8

For Your Safety

Strictly follow the Instructions for Use 

Any use of the apparatus requires a full understanding 
and strict observation of these instructions. The 
apparatus is only to be used for purposes specified 
here. 

Maintenance 

The apparatus must be inspected and serviced 
regularly by suitably qualified persons.
Repairs to the apparatus may only be carried out by 
trained service personnel.
We recommend that you enter into a service contract 
with Dräger Safety and that all repairs be conducted by 
Dräger Safety.
Only authentic Dräger spare parts may be used for 
maintenance.

Accessories

Do not use accessory parts other than those listed in 
the order list.

Use in areas subject to explosion hazards

Equipment or components which have been inspected 
and approved in accordance with national regulations 
concerning electrical equipment in explosion-hazard 
areas (ElexV) or European explosion-protection 
regulations, may be used only under the conditions 
explicitly specified in the approval. 
The equipment or components may not be modified in 
any manner. The use of faulty or incomplete parts is 
forbidden. 
The appropriate regulations must be observed (e.g. 
EIExV and EN 50014) when carrying out repairs on the 
equipment or components.

Liability for proper function or damage 

The liability for the proper function of the apparatus is 
irrevocably transferred to the owner or operator, to the 
extent that the apparatus was serviced or repaired by 
personnel not employed or authorized by Dräger Safety 
or when the apparatus was used in a manner not 
conforming to its intended use.
Dräger Safety cannot be held responsible for damage 
caused by non-compliance with the recommendations 
given above.
The warranty and liability provisions of the terms of sale 
and delivery of Dräger Safety are likewise not modified 
by the recommendations given above.
Dräger Safety AG & Co. KGaA 

Intended Use

Gas Detector Pump for all Dräger short term tubes and 
sampling tubes.

What is What

1

Tube opener with ceramic cutter

2

Cover plate

3

Socket

4

Rubber bellows

5

Lower pump plate

6

Stroke Counter Reset Button

7

Stroke counter

8

End-of-Stroke Indicator (End-of-Stroke membrane)

9

Bolt to secure lower pump plate in place

10 Hole to break off the tube tip
11 Hole for attachment of loop handle
12 Loop handle with snap hook

D

accuro

00

1

2
3

4

5

6

10

7

8

11
12

9

00

1

22003.eps

Preparation

Pump leak test (A)

Insert unopened Dräger tube into socket 

(A1)

.

Squeeze pump 

completely

 and release 

(A2)

.

Pump is adequately leak-proof if the end-of-stroke 
indicator 

(B) 

has not appeared after 15 minutes.

If it has, then see chapter "Fault, Cause, Remedy". 

Remove unopened Dräger tube from socket.

Press button to reset counter to zero 

(C)

button is recessed to prevent accidental resetting.

Preparing the tube (D)

Select appropriate Dräger tube for the measurement 
required. Follow the relevant Instructions for Use!

Open tip with Dräger tube opener 
(item no. 64 01 200).

Or:

Insert the tip into the opening of the lower pump plate 

(D1)

; keep it pressed against ceramic edge while 

turning it, so that the glass is scored.

Pull the tube over at an angle until the tip breaks off.

Discard broken tip.

Open the other end of tube in the same way.

Or:

Turn handle of pump to the side and insert tube tip in 
the free hole of the plate 

(D2)

.

Tilt tube sideways until the tip breaks off. 

Open other end of tube in the same way.

Be careful when opening the Dräger tube as glass 
splinters may come off! Keep the pump facing 
downwards by holding the loop handle, so to prevent 
glass splinters from entering the pump!

Insert Dräger tube (now open at both ends) into 
pump, the arrow must point towards the pump 

(E).

Caution: the tips of the tube may have sharp edges; 
risk of injury!

English

Содержание accuro

Страница 1: ...i utasítás 50 es Instrucciones de uso 12 lv Lietošanas instrukcija 32 el Οδηγίες χρήσης 52 pt Instruções para uso 14 pl Instrukcja obsługi 34 tr Kullanma talimatlar 54 it Istruzioni per I uso 16 ru Пpавила пользования 36 ar 56 nl Gebruiksaanwijzing 18 hr Upute za uporabu 38 zh 58 da Brugsanvisning 20 sl Navodila za uporabo 40 ja 60 fi Käyttöohje 22 sk Návod na použitie 42 ko 62 no Bruksanvisning 24...

Страница 2: ...2 ...

Страница 3: ...3 ...

Страница 4: ...4 accuro C accuro 0 0 A 1 2 B 00221003 eps ...

Страница 5: ...5 I G 2 2 1 H 1 2 1 2 J K E F D 1 2 2 0 0 1 accuro 00321003 eps ...

Страница 6: ...nweise nicht erweitert Dräger Safety AG Co KGaA Verwendungszweck Gasspürpumpe für alle Dräger Kurzzeit Röhrchen und Probenahme Röhrchen Was ist was 1 Keramamikscheibe zum Öffnen des Röhrchens 2 Gehäuseplatte 3 Tülle 4 Gummibalg 5 Untere Griffplatte 6 Taste zum Rücksetzen des Zählers 7 Zählwerk für Anzahl der Hübe 8 Indikator für Hubende Hubende Membran 9 Schraube zum Befestigen der unteren Griffpl...

Страница 7: ...nehmen I2 Die Gehäuseplatte hält die Teile des Indikators für das Hubende in ihrer Position Federnde Stege zusammendrücken und Tülle her ausziehen J Neue Tülle einsetzen Neue Hubende Membran in der abgebildeten Rei henfolge in das Gehäuse einsetzen K Auf richtige Position achten Gehäuseplatte wieder aufsetzen und festdrücken Pumpe auf Dichtheit prüfen A Technische Daten Prüfungen und Zulassungen E...

Страница 8: ...e and delivery of Dräger Safety are likewise not modified by the recommendations given above Dräger Safety AG Co KGaA Intended Use Gas Detector Pump for all Dräger short term tubes and sampling tubes What is What 1 Tube opener with ceramic cutter 2 Cover plate 3 Socket 4 Rubber bellows 5 Lower pump plate 6 Stroke Counter Reset Button 7 Stroke counter 8 End of Stroke Indicator End of Stroke membrane...

Страница 9: ...g ball pen I1 Take out cover plate I2 Cover plate holds End of Stroke Indicator assembly in place Press spring flanges together and remove socket J Insert new socket Using a new End of Stroke Indicator Membrane place parts into the housing in the sequence illustrated K Observe appropriate position Insert spring hook of cover plate and push on firmly Perform the leak test as described A Technical D...

Отзывы: