English / US English
32
Instructions for Use Air Compressor
Dew point depression at operating pressure
5 °C (41 °F) under ambient temperature at
≥
30 L/min
and max. ambient temperature 40 °C (104 °F).
15 °C (59 °F) under ambient temperature at 15 L/min
and max. ambient temperature 40 °C (104 °F)
Air quality
Filtered and oil-free compressed air
Filter
≤
1 µm
Compressed air outlet
Coupling with check valve
High temperature indicator
visual (orange LED) and acoustic (continuous buzzer)
Standby – pressure monitoring
Compressor – compressed air ON
<2.7
±0.2
bar (<40
±3
psi)
Compressor – compressed air OFF
>3.2
±0.2
bar (>46
±3
psi)
Operating characteristics
Connection for central gas supply system
NIST or DISS connection
Compressor Nominal voltage Frequency
Standby mode
Fuses (T1, T2)
5 x 20 mm
Current
consumption
[V]
[Hz]
IEC 127-2/V
[A]
8413417
100
50
yes
T8H250V
5.4
8413894
100
60
yes
T8H250V
5.4
8413418
110
50
yes
T8H250V
4.8
8413899
110
60
yes
T8H250V
4.8
8413419
115-120
60
yes
T8H250V
4.7
8413421
115-120
60
yes
T8H250V
4.7
8413422
127
60
yes
T8H250V
4.2
8413423
127
50
yes
T8H250V
4.2
8413892
220-230
60
yes
T4H250V
2.4
8413897
220-240
50
yes
T4H250V
2.2
8413900
220-240
50
no
T4H250V
2.2
Protection class
I
Electromagnetic compatibility EMC
Tested acc. EN 60601-1-2: 2004 and 2007
Sound pressure level at 50 Hz operation
approx. 49 dB(A) at maximum continuous flow
Sound pressure level at 60 Hz operation
approx. 52 dB(A) at maximum continuous flow
Measured at the front at a distance of 1 m (3.28 ft)
and a height of 1 m (3.28 ft).