7
2.11 Reinstall the drawer on the tracks. Make sure it operates properly without binding.
NOTE:
Proceed to Section 3 or Section 4 if you are installing a display unit at this time.
This completes the NM6000 auxiliary video installation.
2.12 Restore power to the machine and perform the PHASE adjustment on the flat panel
display as given in Section 5.7A of the Narkomed 6000 Technical Service manual.
2.13 Reinstall the table lamp panel at the front of the machine.
2.14 Verify the machine’s self-diagnostics, flat panel display, remote display, aux lamp and
strip chart recorder (if fitted) are working properly.
2.15 Perform electrical safety and battery tests; refer to Section 6 in the Narkomed 6000
Technical Service Manual.
2.16 Locate the Operator’s Manual binder assembly and insert Supplement (4117816),
Auxiliary Video Output Option, behind the supplement tab.
FIGURE 5. Ground Strap and Bracket Installation
SP24405
GROUND STRAP ASSEMBLY
10-32 KEP NUT
GROUND CLAMP
6-32 X 1/4 IN.
BTN HD SKT SCREWS
AND #6 INT-T
LOCKWASHERS
GROUND STRAP
BRACKET
Installation Procedure (continued)
RETURN TO CD-ROM TABLE OF CONTENTS