Dräger Medical AG & Co. KGaA • EvitaXL 

 5

Repair instructions

Replacing the CPU 68332 PCB

7 Remove the module D22 (EEPROM) from the faulty 

CPU 68332 PCB and plug it onto the new CPU 68332 
PCB observing the IC identification, see 

Fig. 3

.

The D22 module contains device data (such as 
device identification, operating hours, enabled 
options, and customer-specific settings). Each device 
has its own device identification. Therefore, the 
module must be transferred from the old CPU 68332 
PCB to the new one. If this module is faulty, the 
device can be operated with the new D22 module 
which is available on each new CPU 68332 PCB. 
However, in this case the device has only the basic 
settings, optional features must be enabled, see 
service diagnosis mode. 

Important: The time must be set!

Fig. 3

CPU PCB 68332

8 Check the jumper configuration and change if 

applicable.

A faulty CPU 68332 PCB is special waste. Dispose of 
a faulty CPU 68332 PCB according to applicable 
waste disposal regulations.