Dräger Medical AG & Co. KGaA • EvitaXL
5
Repair instructions
Replacing the CPU 68332 PCB
7 Remove the module D22 (EEPROM) from the faulty
CPU 68332 PCB and plug it onto the new CPU 68332
PCB observing the IC identification, see
The D22 module contains device data (such as
device identification, operating hours, enabled
options, and customer-specific settings). Each device
has its own device identification. Therefore, the
module must be transferred from the old CPU 68332
PCB to the new one. If this module is faulty, the
device can be operated with the new D22 module
which is available on each new CPU 68332 PCB.
However, in this case the device has only the basic
settings, optional features must be enabled, see
service diagnosis mode.
Important: The time must be set!
Fig. 3
CPU PCB 68332
8 Check the jumper configuration and change if
applicable.
A faulty CPU 68332 PCB is special waste. Dispose of
a faulty CPU 68332 PCB according to applicable
waste disposal regulations.