Dräger Medical AG & Co. KGaA • EvitaXL 

 4

Repair instructions

Replacing the CPU 68332 PCB

Fig. 2

Detail of the CPU 68332 PCB; service LEDs

7

Pneumatics 
disabled

This signal is set if the 
pneumatics is not 
functioning properly. It will 
remain set until the 
pneumatics functions 
properly again. 

LED Reset

Meaning

4

Replacing the CPU 68332 PCB

1 Switch off the device.

2 Unplug the power plug of the device from the socket-

outlet.

3 Create ESD conditions.

4 Remove the faulty CPU 68332 PCB from the devcie 

and place it on the static-dissipative mat.

5 Remove the lithium battery (

Fig. 3

/G2) from the faulty 

CPU 68332 PCB and insert it into the battery holder of 
the new CPU 68332 PCB making sure the polarity is 
correct.

6 Remove the module D22 (EEPROM) (

Fig. 3

/D22) from 

the new CPU 68332 PCB.

Electrostatic discharge may damage electrostatic 
sensitive devices. When handling electrostatic 
sensitive devices, use a static-dissipative mat 
and a static dissipative wrist strap.