Dräger Medical, Inc.

Incubator 8000 IC/SC/NC

Electrical Safety Test with Biotek Model 501

August 2000

Page 6 of 6

  Maximum allowable test values under Single Fault Condition:

Open Ground, Normal Polarity, Normal Neutral:

500 µA

Open Ground, Reverse Polarity, Normal Neural:

500 µA

(According to CAN/CSA - 22.2 No. 601.1 - M90, Sub-clause 19.3, Table IV)

  Switch off Incubator and disconnect power cord from the Biotek 501 Pro and

remove the test equipment

  Close flap below electronics module.