Dräger Medical, Inc.
Incubator 8000 IC/SC/NC
Electrical Safety Test with Biotek Model 501
August 2000
Page 6 of 6
•
Maximum allowable test values under Single Fault Condition:
Open Ground, Normal Polarity, Normal Neutral:
500 µA
Open Ground, Reverse Polarity, Normal Neural:
500 µA
(According to CAN/CSA - 22.2 No. 601.1 - M90, Sub-clause 19.3, Table IV)
•
Switch off Incubator and disconnect power cord from the Biotek 501 Pro and
remove the test equipment
•
Close flap below electronics module.