Dräger Medical, Inc.

Incubator 8000 IC/SC/NC

Electrical Safety Test with Biotek Model 501

August 2000

Page 4 of 6

2.3

Earth Leakage Current (Ground Wire Leakage Testing)

  Remove the red lead with the alligator clip from the Incubator.

  Remove the alligator clip from the red test lead and plug this end into the green

input jack "GROUND" on the back panel on the Biotek 501 Pro.

  Leave all other selections from the previous test the same.

  The Incubator is still switched on.

  Set up the Biotek 501 Pro for the following tests by using the white keys labeled

"Ground", "Neutral" and "Polarity".

  Note: If the corresponding red LEDs for Ground, Neutral, and Polarity are not lit,

they are in the normal position.

  This is a measurement of the leakage current flowing through the ground wire of

the power cord.

  Activate height adjustment (if available) during every following test.

  Maximum allowable test values under Normal Condition:

Open Ground, Normal Polarity, Normal Neutral:

500 µA, but not 0 µA

Open Ground, Reverse Polarity, Normal Neutral:

500 µA, but not 0 µA

(According to CAN/CSA - 22.2 No. 601.1 - M90, Sub-clause 19.3, Table IV)

  Maximum allowable test values under Single Fault Condition:

Open Ground, Normal Polarity, Open Neutral:

1000 µA

Open Ground, Reverse Polarity, Open Neutral:

1000 µA

(According to CAN/CSA - 22.2 No. 601.1 - M90, Sub-clause 19.3, Table IV)