14
15
Toggle Switch
Temperature Compensation Adjustment
compensation mode.
By pressing the Memory/Mute-unmute button to adjust the temperature
from ±0.0 to ±2.0.
button and Measure button for 2-3 seconds to go to the temperature
When the thermometer is on, press both the Memory/Mute-unmute
Note
All the future temperature you are taking will be automatically added the
value you are adjusted.
Turn off Thermometer
The unit will shut down automatically after 10 seconds of no use. Or you
can keep pressing the Measure button for 5-7 seconds.
All memory records will loss when uninstall or reinstall the battery.
All settings will come to default when uninstall the battery. If need
adjust the settings, please power on and make the new settings.
Cau on
The Battery Replacement
Slide down the battery cover along the marked direction. Put two AAA
batteries correctly into the compartment.
Cau on
Remove the batteries if the thermometer will not be used for more than
two months.
7. TIPS TO MAKE CORRECT TEMPERATURE MEASUREMENTS
Record readings twice a day (early morning and late afternoon). Take
the average of the two temperatures to calculate normal oral
equivalent temperature.
A child’s normal temperature can be as high as 99.9°F (37.7) or as low as
97.0°F (36.11). Please note that this unit reads 0.5ºC (0.9°F) lower than a
rectal digital thermometer.
Holding the thermometer for too long in the hand before taking a
measurement can cause the device to warm up. This means the
measurement could be incorrect.
Before placing the thermometer sensor onto the forehead, remove dirt,
hair, or sweat from the forehead area.
It is important to know everyone’s normal temperature when they are
well. This is the only way to accurately diagnose a fever.
Always take the temperature in the same location, since the
temperature readings may vary from different locations on the
forehead.
Patients and the thermometer should stay in steady-state room
condition for at least 30 minutes.
Use an alcohol swab to carefully clean the sensor and wait for 5 minutes
before taking a measurement on another patient.
Wiping the forehead with a warm or cool cloth may impact your
reading. It is advised to wait for 10 minutes before taking a reading.
8. MULTI MEASUREMENTS REQUIREMENTS
In the following situations 3-5 measurements recommended to be taken in
the same location. The highest one is the standard reading.