Страница 5: ...0 and80 Self check NormalBP between120 and 130 between80 and 85 Self check High To NormalBP between130 and 140 between85and 90 Consult your doctor Slight Hypertension between140 and 160 Between90and 100 Seek medical advice Medium Hypertension between160 and 180 Between100and 110 Seek medical advice Strong Hypertension Higher than 180 Higher than 110 Urgently seek medical advice ...

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ...urement Artefact ERR 3 The inflation of the cuff takes too long The cuff is not correctly seated ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures Take other reading following directions carefully Contact you doctor if you continue to get unusual readings Err8 The pressure in the cuff is too high over 290 mmHg ...

Страница 10: ...tails made under point 5 Every measurement produces a different value although the instrument functions normally and the values displayed are normal 1 Please read the following information and the points listed under Common sources of error Repeat the measurement Please Note Blood pressure fluctuates continually so successive measurements will show some variability Blood pressuremeasured differs f...

Страница 11: ...N60601 1 2 2015 IEC EN60601 1 11 2015 IEC80601 2 30 2009 A1 2013 EN80601 2 30 2010 A1 2015 Some electrical and electrical equipments forbid to abandon and disposal at will TUV NO Manufacturer s name and address Reading Instruction Book before use Inapplicable baby Type B equipment Cuff Connector Protection against ingress of water IP22 ...

Страница 12: ......

Страница 13: ...nterruptions it is recommended that Dr Trust BP Check with MDI be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Not applicable Not applicable Note U T is the a c mains voltage prior to application of the test level Immunity test IEC60601 1 2 test level IEC60601 1 2 test level Electromagnetic environment guidance Conducted RF I...

Страница 14: ...mitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 p1 2 80 MHz to 800 MHz d 1 2 p1 2 800 MHz to 2 5 GHz d 2 3 p1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 ...