Immunity test   

IEC60601-1-2 test 
level   

Compliance level   

Electromagnetic environment 
-guidance   

Electrostatic 
discharge (ESD) IEC 
61000-4-2   

±6 kV contact 
±8 kV air   

±6 kV contact 
±8 kV air   

Floors should be wood, concrete 
or ceramic tile. If floors are covered 
with synthetic material, the relative 
humidity should be at least 30 %.   

Electric fast 
transient/ burst IEC 
61000-4-4   

±2 kV for power 
supply lines 
±1 kV for 
input/output lines   

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment.   

Surge IEC 61000-4-5   

±1 kV differential 
mode   
±2 kV common 
mode   

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment.   

Voltage dips, short 
interruptions and 
voltage variations on 
power supply input 
lines IEC 61000-4-11   

<5 % U

T

95% dip 

inU

T

.

  for 0.5 cycle   

40 % U

T

60% dip in 

U

T

 

for 5 cycles 

70 % U

T

(30% dip 

inU

T

) for 25 cycles 

<5 % U

T

95% dip 

inU

T

for 5 sec. 

Not applicable 

Mains power quality should be 
that of a typical commercial or 
hospital environment. If the user of 
the upper arm style requires 
continued operation during power 
mains interruptions, it is 
recommended that Dr Trust BP 
Check with MDI be powered from 
an uninterruptible power supply or 
a battery.   

Power frequency   
(50/ 60 Hz) magnetic 
field IEC 61000-4-8   

3 A/m   

Not applicable 

Not applicable 

Note: U

T

 is the a.c. mains voltage prior to application of the test level. 

 

Immunity test  

IEC60601-1-2 
test level  

IEC60601-1-2 
test level  

Electromagnetic environment 

- guidance 

 
 
 
 
 
 
 

Conducted RF 

IEC 61000-4-6 

 
 

Radiated RF 

IEC 61000-4-3 

 
 
 
 
 
 
 

3 Vrms 150 kHz 
to 80 MHz 80% 

AM (2Hz) 

 
 

3 Vrms 80 MHz 

to 2.5 GHz 80% 

AM (2Hz) 

 
 
 
 
 
 
 
 

3 Vrms 

 
 
 

3 V/m 

Portable and mobile RF communications equipment 
should be used no closer to any part of Dr Trust BP Check 
with MDI , including cables, than the recommended 
separation distance calculated from the equation 
applicable to the frequency of the transmitter. 

 

Recommend separation distance 

 

3V 
d = 1.2×p

1/2

    80Mhz to 800 MHz 

 

d = 2.3×p

1/2

MHz to 2.5 GHz  

Where 

is the maximum o

utput power rating of the 

transmitter in watts (W) according to he transmitter 
manufacturer and d is the recom

mended separation 

distance in meters (m).

 

Field strengths from fixed RF transmitters as determined 
by an electromagnetic site survey

a

, should be 

less than the 

compliance level in each frequency range

b

. Interference 

may occur in the vicinity of equipment marked with the 
following symbol: 

 

 

 

Содержание Comfort

Страница 1: ......

Страница 2: ......

Страница 3: ......

Страница 4: ......

Страница 5: ...0 and80 Self check NormalBP between120 and 130 between80 and 85 Self check High To NormalBP between130 and 140 between85and 90 Consult your doctor Slight Hypertension between140 and 160 Between90and 100 Seek medical advice Medium Hypertension between160 and 180 Between100and 110 Seek medical advice Strong Hypertension Higher than 180 Higher than 110 Urgently seek medical advice ...

Страница 6: ......

Страница 7: ......

Страница 8: ......

Страница 9: ...urement Artefact ERR 3 The inflation of the cuff takes too long The cuff is not correctly seated ERR 5 The measured readings indicated an unacceptable difference between systolic and diastolic pressures Take other reading following directions carefully Contact you doctor if you continue to get unusual readings Err8 The pressure in the cuff is too high over 290 mmHg ...

Страница 10: ...tails made under point 5 Every measurement produces a different value although the instrument functions normally and the values displayed are normal 1 Please read the following information and the points listed under Common sources of error Repeat the measurement Please Note Blood pressure fluctuates continually so successive measurements will show some variability Blood pressuremeasured differs f...

Страница 11: ...N60601 1 2 2015 IEC EN60601 1 11 2015 IEC80601 2 30 2009 A1 2013 EN80601 2 30 2010 A1 2015 Some electrical and electrical equipments forbid to abandon and disposal at will TUV NO Manufacturer s name and address Reading Instruction Book before use Inapplicable baby Type B equipment Cuff Connector Protection against ingress of water IP22 ...

Страница 12: ......

Страница 13: ...nterruptions it is recommended that Dr Trust BP Check with MDI be powered from an uninterruptible power supply or a battery Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m Not applicable Not applicable Note U T is the a c mains voltage prior to application of the test level Immunity test IEC60601 1 2 test level IEC60601 1 2 test level Electromagnetic environment guidance Conducted RF I...

Страница 14: ...mitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 p1 2 80 MHz to 800 MHz d 1 2 p1 2 800 MHz to 2 5 GHz d 2 3 p1 2 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 ...

Отзывы: