EMC Tables
120
is intended for use in the electromagnetic environment specified below. The customer or the
user of
120
must make sure that it is used in such an environment.
Guidance and manufacturer’s declaration - Electromagnetic emissions
Phenomenon
Professional healthcare facility
environment a)
HOME HEALTHCARE
ENVIRONMENT a)
Conducted and radiated
RF EMISSIONS
a)
CISPR 11
Group 1 Class B
Harmonic distortion
Not applicable
Voltage fluctuations and
flickering
Not applicable
a) The equipment is suitable for use in Home Health Environments and Professional Health Care
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics.
The more restrictive acceptance limits of Group 1 Class B (CISPR 11) have been considered
and applied. The equipment is suitable for use in the mentioned environments when directly
connected to the Public Mains Network.
b) The test is not applicable in this environment unless the ME EQUIPMENT and ME SYSTEM
used will be connected to the PUBLIC MAINS NETWORK and the power input is otherwise
within the scope of the Basic EMC standard.
Guidance and manufacturer’s declaration - Electromagnetic immunity - Enclosure port
Phenomenon
Basic EMC
standard or test
method
mmunity test levels
Professional
healthcare facility
environment
HOME HEALTHCARE
ENVIRONMENT
ELECTROSTATIC
DISCHARGE
IEC 61000-4-2
± 8kV contact
± 2 kV, ±4kV ±, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
a)
10 V/m b) 80MHz - 2.7
GHz 80% AM at 1kHz
Proximity fields from RF
wireless communications
equipment
IEC 61000-4-3
COMPLIANT
NOTE: Further information about distances to
be maintained between portable and mobile RF
communications equipment (transmitters) and
the 120 can be requested from using the contact
information provided in this manual. However,
it is advisable to keep the electromechanical
aerosol equipment at an adequate distance of,
at least, 0.5 m from mobile phones or other RF
communications transmitters to minimise possible
interference.
RATED power frequency
magnetic fields.
IEC 61000-4-8
30 A/m c)
50 Hz or 60 Hz
a) The equipment is suitable for use in Home Health Environments and Professional Health Care
Environments limited to patient rooms and respiratory treatment facilities in hospital or clinics.
The more restrictive IMMUNITY acceptance limits have been considered and applied.
b) Before modulation is applied.
c) This test level assumes a minimum distance of at least 15 cm between the ME EQUIPMENT
or ME SYSTEM and sources of power frequency magnetic fields.
15
www.drtrust.in