Mach M2
Dr. Mach
Lamps and Engineering
Sterilisation
Only previously cleaned and disinfected handles may be sterilised.
The handles are placed in a suitable sterilisation pack (one-way sterilisation pack, e.g. foil / paper sterili-
sation bags, single or double pack) in accordance with DIN EN 868 / ISO 11607 for steam sterilisation
and then sterilised.
Use only the sterilisation procedure listed below for sterilisation. Other sterilisation procedures (e.g. ethyl-
ene oxide, formaldehyde and low-temperature plasma sterilisation) are not permissible.
Steam sterilisation procedure
Validated in accordance with DIN EN 554/ISO 11134
Maximum sterilisation temperature 134°C
Proof in principle of the handles’ suitability for effective sterilisation was provided using a fractional vac-
uum process (Euroselectomat 666 by MMM Münchner Medizin Mechanik GmbH, sterilising temperature
134°C, holding time 7 min.)
Inspection / durability
The handles should be inspected for damage and changed before re-use, if required.
The handles may be cleaned / disinfected, sterilised and re-used for a maximum of 1000 times. If the
handles are re-used more than 1000 times, then this will be the responsibility of the hospital / clinic.
3.2 Lamp head, splinter protection disk
The lamp has a high-quality surface. The surface
can be cleaned with conventional cleaning agents.
The
splinter protection disk
5
is made of a high-
quality plastic. Pay attention to the following during
cleaning:
•
Always clean the
splinter protection disk
5
with a wet cloth (never clean with a dry cloth!).
•
Only use disinfectants with
less than 20% al-
cohol
.
Wipe the
splinter protection disk 5
after cleaning
with an antistatic, non-fluffy cloth.
5
Alc.
≤
20 %
59120001
Edition 04
07.03.2005 / Bak page 8/22