3. Cleaning
3.1 Handle
The sterilizable handle
1
can be removed and steril-
ised at temperatures up to a max. of 134°C.
No other parts may touch the handle during sterilisa-
tion.
The handle should be placed in a vertical position
during sterilisation.
1
3.2 Lamp housing, dispersing lens
and support system
The MACH OT-lamp system has a high-quality sur-
face, which can be cleaned with conventional clean-
ing agents.
The lens system
2
is made of a high-quality plastic.
Pay attention to the following during cleaning:
-
Never wipe over the lens system
2
with a dry
cloth.
-
Only use disinfectant with less than 20% alco-
hol.
Wipe the lens system
2
after cleaning with an anti-
static, non-fluffy cloth.
2
Alc.
≤
20 %
59210001
Edition 04
10.12.2001 / Bak Page 17/30