Dr. A. Kuntze K 100 Cl2 Скачать руководство пользователя страница 23

23

K 100 (W) Cl

2

, SCl

2

, ClO

2

, O

2

, O

3

, H

2

O

2

.

.

Averaging

.

.

Cleaning

.

.

Basic settings

Main menu

Basic settings

Cleaning/day

         1

Turn-on delay

       6.0 h

Cleaning

5.3 Automatic sensor cleaning ASR (option)

5.3 Automatic sensor cleaning ASR (option)

The patented Automatic Sensor Cleaning ASR prevents staining and passivating
coatings and keeps the surface of metal sensors clean throughout the measurement.

The cleaning is an electrochemical one: in adjustable intervals the instrument applies
a strong voltage to the electrode to produce Hydrogen and Oxygen from the
surrounding water. This method needs no addition of chemicals.

The cleaning process takes about 30 seconds. During this time measurement is not
possible, and after cleaning the sensor needs some time for polarisation. Therefor the
display and the analog output will show the last measured value for five minutes, and
the message „cleaning in progress“ is displayed. As a safety measure, attempts to
calibrate within these five minutes are ignored.

Activation and timing

The cleaning is activated by setting the cleaning function from 0/day (never) to 1/day
(every 24h) or 2/day (every 12 h). The first cleaning starts as soon as the function is
activated, and subsequent cleanings are carried out after 24h, or 12h, according to
settings.

The start delay allows to select a more convenient cleaning time, e.g. in the night. After
activation the cleaning is delayed by the set time.

NOTE

Whenever the instrument indicates „no water“, cleaning is not carried
out.

5.4  Averaging

In the basic settings you can activate an averaging function to smoothen the signals.
This is especially useful when using membrane sensors.

Averaging

ON

Averaging

Содержание K 100 Cl2

Страница 1: ...100 Cl2 SCl2 ClO2 O2 O3 H2 O2 K 100 W Cl2 SCl2 ClO2 O2 O3 H2 O2 Measuring and control instruments for potentiostatic measurements of Chlorine Total Chlorine Chlorine dioxide Oxygen Ozone and Hydrogen...

Страница 2: ...Dr A Kuntze GmbH Robert Bosch Str 7a D 40668 Meerbusch Tel 49 21 50 70 66 0 Fax 49 21 50 70 66 60 info kuntze com www kuntze com Dr A Kuntze Gutes Wasser mit System...

Страница 3: ...n 21 5 2 Temperature compensation 22 5 3 Automatic sensor cleaningASR option 23 5 4Averaging 23 6 Adjustment of the controller 24 6 1 ON OFF controller 25 6 2 P PI controller as impulse frequency cont...

Страница 4: ...ed controller which provides bidirectional PI control via two relays or as a steady state controller via the analog output You can define a turn on delay to prevent incorrect dosage after power failur...

Страница 5: ...ment Keep this manual at a safe place where you can always look up the safety instructions and the informations on handling and usage According to DIN 61010 the manual is part of the product and must...

Страница 6: ...ched off until the failure has been taken care off CAUTION The instrument checks the input signals the calibration results and the water flow if a flow sensor is connected It cannot detect erroneous s...

Страница 7: ...direction Controller types ON OFF controller with hysteresis P controller as Pulse Pause Impulse Frequency or steady controller PI controller as Pulse Pause Impulse Frequency or steady controller Hys...

Страница 8: ...bles up to 1 5mm2 for cables up to 1 5mm2 Protection class Front IP54 IP65 Power supply 230 V 6 10 40 60 Hz optionally 117 V or 24 V internal fuse none 230V 63mA HRC 117V 125mV HRC 24V 800mA NRC Power...

Страница 9: ...a place where it is not put under mechanical or chemical strain Mind the protection class K 100 Front IP54 K 100 W IP65 closed terminal cover Connections You will find detailed connection diagrams on...

Страница 10: ...K 100 W Cl2 SCl2 ClO2 O2 O3 H2 O2 10 2 1 Dimensions K 100 W K 100 2 1 Dimensions...

Страница 11: ...Installation K 100 Prepare an opening of 92 x 92 mm Screw tight until the instrument is fixed perfectly Install the instrument from the front side and fix it with the two mounting clips which were par...

Страница 12: ...ways for installation 1 You can hang the instrument upon the upper screw In that case drill the upper hole 120mm above the lower two 2 Or you can slip the fixture on the back of the isntrument under...

Страница 13: ...1 Screen electronics 2 Measuring electrode 3 reference electrode 4 counter electrode With 2 electrode sensors short circuit terminals 3 4 Membrane sensor 1 4 1 2 measurement with electronics 3 6 V 4 6...

Страница 14: ...trode 3 Reference electrode 4 Counter electrode Membrane sensors 1 4 1 reference GND with electronics 2 Measuring signal 3 6 V 4 6 V Pt100 NTC 6 7 Display Kontrast Display Potentiometer to adjust brig...

Страница 15: ...and the status of the relays S1 and S2 both OFF With five membrane keys you can move within the menu With key you enter the main menu With keys and you move up and down in the menu With key you adress...

Страница 16: ...k to where you started or until you reach the alternative you were looking for With these parameters any changes are immediately valid there is no need to store the change 1 Adress the parameter with...

Страница 17: ...ation Data output Code and language Data output Adjustment of the meter averaging Adjustment of the meter temperature compensation Adjustment of the meter automatic Sensor cleaning Adjustment of the c...

Страница 18: ...lse freq S2 00 100 h Direction S1 raise Direction S2 lower Hysteresis 0 01 mg l Pulse Pause 10 sek Min pulse 0 1 sec Dosage check S1 060min Dosage check S2 060min Start delay ON Start delay 180 sec Se...

Страница 19: ...11 gives access to the parameters calibration temperature compensation and set points Code 86 gives access to all parameters and functions With any other number it is impossible to select view or cha...

Страница 20: ...e actual concentration of the test water is determined by an alternative method for example photometrically with DPD and this value is entered as calibration value Calibration of the Oxygen measuremen...

Страница 21: ...he calibration is interpreted as zero point correction and the instruments substracts this value from all subsequent measurements However zero point correction is only advisable if Analog input 1 in t...

Страница 22: ...ion better switch to manual compensation 2 Manual compensation If the temperature is more or less stable you can enter it manually The instrument will then compensate the temperature effect of this te...

Страница 23: ...s some time for polarisation Therefor the display and the analog output will show the last measured value for five minutes and the message cleaning in progress is displayed As a safety measure attempt...

Страница 24: ...s the dosage in the vicinity of the set point proportional to the control deviation This is easily achieved if the analog output is used as steady control output If the relays are used the proportiona...

Страница 25: ...ng parameters 1 Set points S1 and S2 Set point S1 refers to relay 1 set point S2 refers to relay 2 2 P range and integral action time for S1 and S2 For an ON OFF controller set P range 0 and integral...

Страница 26: ...r as impulse frequency controller You can choose different controller versions for S1 and S2 For an impulse frequency controller you have to set the following parameters 1 set points S1 and S2 S1 refe...

Страница 27: ...e For a pulse pause controller you have to set the following parameters 1 set points S1 and S2 S1 refers to relay 1 S2 refers to relay 2 2 P range and integral action time Adjust a P range 0 For a P c...

Страница 28: ...n delay Set a delay time which has to pass before the controller is activated after start up or power interrupt This allows the measurement to settle and prevents inappropriate dosage of chemicals 6 6...

Страница 29: ...ight 5 Switch to the operation mode of S2 with key The square to the left of S2 starts to flash 6 Switch ON relay 2 with key The square to the left of S2 gets dark 7 Switch OFF relay 2 again with key...

Страница 30: ...ler output is 100 for more than the specified time this is interpreted as an indication of failure and the instrument issues an alarm and deactivates the controller thus stopping further dosage The do...

Страница 31: ...ons are immediately evident With the parameters Begin and End you define which part of the measuring range you want to read out 7 2 Current output as controller output Alternatively you can use the cu...

Страница 32: ...message limit 1 or 2 respectively and relay 3 is switched ON This relay can be used to activate an external horn or lamp NEW You decide whether the limit function is active only when the controller is...

Страница 33: ...s not receive a correct signal an alarm is issued and the controller deactivated Alarm and controller stop remain until the analog input receives correct signals again In case of an input error the cu...

Страница 34: ...NC In that case the relay is kept closed actively so in case of power failure it opens thus issuing an alarm NEW You can also decide whether the relay should switch permanently during an alarm or whe...

Страница 35: ...riod of time The digital input has been short circuited The Automatic Sensor Cleaning has started less than five minutes ago Please check the connections the cable and the sensor for signs of damage T...

Страница 36: ...l data To exchange the fuse open the front carefully The fuse is located in the lower right hand side It is kept in place by a Bayonet lock Turn the lock to the left until the fuse pops up Exchange it...

Страница 37: ...me time to repolarise after cleaning For maintenance of membrane sensors please refer to the instructions in the sensor manual Recalibrate in regular intervals NOTE The instrument checks the calibrati...

Страница 38: ...nstrument hardware and software Analog inputs Here you can see the raw data the instrument obtains from the sensors They are not influenced by compensations or calibration and offer valuable informati...

Страница 39: ...m K 100 W 14 Connection diagram K100 13 Connections 9 Controller Acting direction 25 ON OFF controller 24 P controller 24 PI controller 24 Controller output 31 Controller settings Impulse frequency co...

Страница 40: ...Menu Basic settings 17 Main menu 17 Menuoverview 18 P Parameter setting 16 Password 19 Product info 38 Protection class 8 Pulse contact 34 R RC filter 9 Registration output 31 Relay contact rating 8...

Страница 41: ...controller S1 controller S2 Begin End Controller Controller S1 Controller S2 Direction raise lower Direction raise lower Set point Set point Hysteresys Hysteresys P range P range Integral time sec Int...

Отзывы: