15
K 100 (W) Cl
2
, SCl
2
, ClO
2
, O
2
, O
3
, H
2
O
2
1 2 3 4 5 6
0.49 mg/l 25.0°C
S1
S2 Man
7 8 9 10 11
1
Measured value
2
Status relay 1
3
Status relay 2
relay OFF
relay ON
4
Temperature
5
Controller
AUTO: controller ON
MAN : controller OFF
(manual operation
of the relays)
HOLD: freezes analog
output and limit function
6
Orientation aids
7
Key left (
)
8
Key up (
)
9
Key down (
)
10
Key right (
)
3. Operation of the instrument
3. Operation of the instrument
When turned on the instrument shows the measured value and temperature together
with the controller mode (Man) and the status of the relays S1 and S2 (both OFF).
With five membrane keys you can move within the menu:
With key
you enter the main menu.
With keys
and
you move up and down in the menu.
With key
you adress a menu or parameter.
With key
you leave a menu or store a change.
For your convenience triangles in the display indicate the directions you can take from
your position in the menu.
From the display of the measured values you can switch the controller ON and OFF
with key
in the order: AUTO
HOLD
MAN
AUTO.
With key „M“ you get back to the display of the measured values from any point in the
menu. Changes that have not been stored will be lost.
NOTE
The instruments K 100 and K 100 W differ only in housing. Operation
and menus are the same.