Square Sequence(s)
(a)
Square Sequence(s) Mode Interface
(b)
Exemple of Error
Figure 4.8:
Sequence Options of the Square Sequence(s) Mode.
The
Square Sequence(s)
mode allows the creation of a square TTL pulse sequence. The Sequence(s) Options of this
mode are shown in Figure
and are explained below.
1. The
Starting Delay
defines the time between the activation of the pulse sequence and the beginning of the first
light illumination.
2. The
Frequency
sets the frequency (in Hz), which is the number of pulses per second. The frequency can also be
changed to the
Period
. For example, a light illumination at 10 Hz (frequency) will output one pulse every 100
ms (period), whereas a light illumination at 0.5 Hz (frequency) will output one pulse every 2 seconds (period).
3. The
Time ON
defines the length of a single pulse. This time can also be converted to a
Duty Cycle
, which
represents the % of the period the pulse duration corresponds to.
4. The
Smoothing
check box allows to change the pulse slope in square pulse sequences. The
Edit Edges
button
opens the
Smoothing Edge(s)
window. An overview of the window opened by
Edit Edges
will be done in the
next subsection.
5. The
Pulse(s) per sequence
set the number of pulses per sequence. If it is set to 0, the number of pulses will be
infinite.
6. The
Number of sequence(s)
sets the number of times that the sequence will be repeated.
7. The
Delay between sequences
sets the delay between each sequence.
8. The
Total Duration
shows the total expected duration of the pulse sequence. Should the duration be infinite,
the box will display
∞
. If there is an error in parameter selection, this box will turn red and display what is the
error (see Figure
Chapter 4. Doric Neuroscience Studio
18