Figure 3.5:

Driver and Light Source in External Analog Mode

Figure 3.6:

Light Source Channel Configuration Window, square sequence options

d) The

Square sequences

mode has the light source follow a square pulse sequence.

i. The

Starting Delay

(Fig.

3.6

sets the delay (in hh:mm:ss:zzz format) before the first pulse.

ii. The

Frequency/Period

(Fig.

3.6

sets the frequency (in Hz) or period (in ms) for the pulse sequence.

For example, a signal at 10 Hz (frequency) will output one pulse every 100 ms (period), whereas a pulse

sequence at 0.5 Hz (frequency) will output one pulse every 2000 ms (period).

iii. The

Time ON/Duty Cycle

(Fig.

3.6

sets the time (in ms) or the duty cycle (in %) for each pulse. The

Time

ON

must be lower than (1/fre0.005 ms, while the

Duty cycle

must be below 100 %. These

squares will appear red should an impossible

Frequency

/

time ON

be selected. Should the

Smoothing

option be selected, this feature becomes inaccessible.

Chapter 3. Doric Neuroscience Studio

13

Содержание LEDFRJ Series

Страница 1: ...LEDs with Fiber optic Rotary Joint User Manual Version 2 1 2...

Страница 2: ...uide 6 2 1 Getting Started 6 2 2 FC Connector Installation 7 2 3 Stand alone mode without Doric Neuroscience Studio Software 8 2 4 Connected to Doric Neuroscience Studio Software 9 3 Doric Neuroscienc...

Страница 3: ...Joint a Standard mount b Rotor Yellow and stator black Figure 1 1 Connectorized LED with Fiber optic Rotary Joint The Connectorized LED with Fiber optic Rotary Joint is a rotary joint with a single in...

Страница 4: ...r Black with the light leaving through a Rotor Yellow Fig 1 3 They contain the following elements The M8 connector ports are used to provide power to the LEDs There is a single port per LED The pinout...

Страница 5: ...with any driver having at least as many channels as the number of combined LEDs For more details on driver specifications and operation see the LED Driver Manual Figure 1 4 1 2 and 4 channel LED Drive...

Страница 6: ...talled in its holder Holder FRJ large GH FRJ It can then be mounted into the setup using 1 4 or M6 nuts and screws If using a Combined LED model the holder is already integrated It can then be mounted...

Страница 7: ...limits indicated in table 4 2 On the Combined LED models it is critical not to block or insert objects into the grids as this could block airflow and reduce cooling efficiency 10 When not in use place...

Страница 8: ...t The operating mode and the maximum driving current setting are independently adjusted for each channel Constant current CW Figure 2 3 Constant Current Mode Driver Signal When using the CW mode the u...

Страница 9: ...oric Neuroscience Studio Software 1 Connect the LED Driver to the power outlet with the included 12 V power adapter and turn the LED Driver switch ON Always power the LED Driver before connecting the...

Страница 10: ...rolling a light source of its given type These channels accessible using the Add Channel will be the first detailed Figure 3 1 Light Source Driver Tab 3 1 1 Channels Each light source driver is separa...

Страница 11: ...CAN OTHERWISE DAMAGE THE LIGHT SOURCE When using a LEDD the Low Power checkbox will appear When selected this allows reduced power signaling for the same voltage This mode is only available for CLED m...

Страница 12: ...e triggered manually When this mode is active the driver channel will show CW under MODE This mode has no additional sequence options Figure 3 4 Driver Signal Response to External Source in External T...

Страница 13: ...n ms for the pulse sequence For example a signal at 10 Hz frequency will output one pulse every 100 ms period whereas a pulse sequence at 0 5 Hz frequency will output one pulse every 2000 ms period ii...

Страница 14: ...ile the Smoothing option is active the Time ON is fixed at this value v The Pulses per sequence Fig 3 6 sets the number of pulses per sequence If it is set to 0 the pulse will be repeated indefinitely...

Страница 15: ...he other will adjust automatically The Time ON must be lower than 1 frequency 0 005 ms while the Duty cycle must be below 100 Figure 3 9 Internal Complex Mode Pulse Sequences viii The Types of pulses...

Страница 16: ...re 3 12 Control Settings Time Series Window The Number of series Fig 3 12 sets the number of times that the sequence will be repeated with a minimum of 1 The Time Active sets the duration of each seri...

Страница 17: ...View Light Source Channel 1 The Light Source Channel box Fig 3 13 contains all elements related to a single light source channel 2 The Scope Channel box Fig 3 14 is used to control and configure an ac...

Страница 18: ...in kilosamples per second at which mea surements are taken j The Start scope channel button activates a live measurement sequence Important measurements should not be made as a live measurement as th...

Страница 19: ...0 63 NA 365 12 6 0 23 100 1 385 12 6 0 23 100 1 405 15 5 0 23 100 1 420 15 5 5 23 100 1 450 25 8 0 23 100 x1 7 465 25 7 5 23 100 x1 7 505 30 3 0 12 50 x1 6 515 40 3 0 9 5 40 x1 5 560 100 2 0 8 5 40 5...

Страница 20: ...Torque Multiple LED 30 N m Electrical connector M8 4 pins male Dimensions Single LED model 70 x 39 x 39 mm3 without holder 2 LED model 62 x 84 x 119 mm3 including holder 3 and 4 LED model 62 x 93 x 14...