Table 3.4:

General Specifications

SPECIFICATIONS

VALUE

NOTES

HDMI connector pinout type

Microscope, Blackrock 2

Female connector

Harwin connector type

Datamate L-Tek serie

Male connector, 2 mm pitch, 12 contacts, 2 rows

Number of contacts

6 or 12

-

Contact material

Gold

-

Maximum current

2 A per contact

-

Contact resistance

<

500 m

-

Resistance variation during rotation

<

100 m

@ 5 VDC

During constant rotation

Start up torque

0.9 mN

·

m (for 6 contacts)

Typical Value

1.9 mN

·

m (for 12 contacts) Typical Value

Rotation speed

up to 300 rpm

-

Outer diameter

45.0 mm

-

Through-hole diameter

7.5 mm

-

Length

59.4 mm

-

Mass

123.0 g

-

Table 3.5:

Recommended Environmental Specifications

DESCRIPTION OPERATION

STORAGE

Use

Indoor

Indoor

Temperature

0-40

C

0-40

C

Humidity

40-60% RH, non condensing 40-60% RH, non condensing

Chapter 3. Specifications

13

Содержание Electrical Rotary Joint

Страница 1: ...Electrical Rotary Joint User Manual Version 1 1 1...

Страница 2: ...lectrical Rotary Joint Holders 4 1 3 Rotary Joint Harwin 12 Omnetics PZN12 Adapter Kit 5 2 Operations Guide 6 2 1 Rotary Joint Holders 6 2 2 Electrical Optical Cables 9 3 Specifications 12 4 Support 1...

Страница 3: ...it signal It is composed of a Stator which stays immo bile and a Rotor that moves Fig 1 1b The electrical signal is transmitted through 2 possible types of connectors Male Harwin 6 or 12 contacts or F...

Страница 4: ...usedincombinationwiththeHolder FRJ smallortheHolder FRJ large If the rotary joint is used only for electric signal the Holder FRJ large alone Fig 1 1 The gimbal mount GH FRJ Fig 1 1 can also be used w...

Страница 5: ...win 12 pin connector on one side and a Omnetics PZN12 connector on the other To use simply insert the Female Harwin connector in the Male Harwin connector on the Stator side of the rotary joint The pi...

Страница 6: ...2 Operations Guide 2 1 Rotary Joint Holders 2 1 1 Holder ERJ a b c d Figure 2 1 Installation of the rotary joints in the Holder ERJ 6...

Страница 7: ...er FRJ small is then be attached to the Holder ERJ frame 1 using 1 4 or M6 screws 2 and bolts 3 Fig 2 1d See Figure 2 2a for the final result For installation using the 1x2 Fiber optic Rotary Joint In...

Страница 8: ...or M6 screws Figure 2 3 Installation of the ERJ in the Holder FRJ large 2 1 3 GH FRJ The Gimbal Holder allows swivel movement of the rotary joint along 2 axes The rotary joint is to be threaded into t...

Страница 9: ...stalling the cables When not in use install plastic caps on connectors for protection and cleanliness 2 Connect the signal electrical cable to the fixed electrical connector Fig 2 5a and 2 5b 3 Connec...

Страница 10: ...a subject patch cord connecting to the experimental subject 1 Connect the signal patch cord to the fixed connector of the optical rotary joint 2 Install the ERJ Holder as shown in section 2 1 1 3 Sele...

Страница 11: ...Figure 2 7 Electrical and Fiber optic Rotary Joints Cable Installation Chapter 2 Operations Guide 11...

Страница 12: ...ctrical Connector Configuration Male on rotary joint Female on adapter Table 3 2 HDMI Electrical Connector Configuration Male on cable Female on rotary joint Table 3 3 HDMI Electrical Connector Pinout...

Страница 13: ...resistance 500 m Resistance variation during rotation 100 m 5 VDC During constant rotation Start up torque 0 9 mN m for 6 contacts Typical Value 1 9 mN m for 12 contacts Typical Value Rotation speed...

Страница 14: ...urn instructions This warranty will not be applicable if the unit is damaged or needs to be repaired as a result of improper use or operation outside the conditions stated in this manual For more info...

Отзывы: