5

Weighing a Patient

1.  Press On/Off.  The scale will turn on and display a zero weight.
2.  Place patient on the scale.
3.  When a stable weight is reached the display will lock and the HOLD

annunciator will be displayed.

4.  The weight will be locked until the scale shuts off.

Zero Operation

1.  Place item to be zeroed out on the scale platform.
2.  Press On/Off.  The scale will turn on and display a zero weight. If the item is

removed, a negative weight will be displayed.

Note: Each time the scale is powered up, the zero point is reset.

3.  Place the patient on the scale and the patient’s weight, less the item’s weight

will be displayed.

OR

1.  With the scale on, press ZERO.
2.  The display will display dashes.
3.  When scale becomes stable, the display will read zero.

Reweigh Operation

With the patient on the scale platform, press REWEIGH to reinitiate the
weighing process.

Содержание DS4100

Страница 1: ...Model DS4100 Infant Scale Instruction Manual Doran Scales Inc 1315 Paramount Pkwy Batavia IL 60510 USA 1 800 264 4107 www doranscales com MAN0259 Revision 0 5...

Страница 2: ......

Страница 3: ...eration 5 Section 3 Scale Calibration 6 Two Point Calibration Procedure 6 Multiple Point Calibration 7 Calibration Troubleshooting 8 Section 4 Parameter Setup 9 Entering and Exiting Setup Mode 9 Setup...

Страница 4: ...oper handling or abuse can damage the scale and result in costly repairs that may not be covered by the warranty Please observe the following precautions to insure years of trouble free service from y...

Страница 5: ...e batteries These batteries supply a year s of typical use with the standard Automatic Shutoff Timer settings The batteries can be accessed on the bottom of the scale To access the battery compartment...

Страница 6: ...hutoff Timer that will power down the indicator when not in use Zero When the zero button is pressed the scale will be reset to zero weight This feature can be used to zero out any weight not desired...

Страница 7: ...scale status Center of Zero Indicates that the scale has an accurate zero reading Hold Indicates a stable weight has been achieved and held on the display This indicates an accurate weight has been a...

Страница 8: ...platform 2 Press On Off The scale will turn on and display a zero weight If the item is removed a negative weight will be displayed Note Each time the scale is powered up the zero point is reset 3 Pl...

Страница 9: ...RO and REWEIGH 5 If desired press LB to use a span calibration of Cal 45 lb CAL20 kg NOTE Using a heavier calibration weight will increase scale accuracy 6 Press KG The display reads Cal 0 7 Remove al...

Страница 10: ...B button until the display reads CALSP 6 Press KG and the display will read CAL 0 7 Remove all weight from the scale NOTE During calibration be sure vibration and air currents are not present 8 Press...

Страница 11: ...AL10 kg release ZERO and REWEIGH 3 Press KG until the RAW A D count is displayed 4 Apply test weights and compare results to the table below 5 Press KG to exit The following table shows the acceptable...

Страница 12: ...exit the setup mode press KG until the display returns to the weigh mode Setup Mode Navigation 1 Press KG to change the current parameter 2 Press LB to change the parameter setting Calibration Weight...

Страница 13: ...10 seconds 30 30 Seconds 60 1 minute 120 2 minutes 180 5 minutes 300 10 minutes 600 15 minutes of Automatic shutoff timer disabled LB Units Button Enable Disable on LB button active off LB button dis...

Страница 14: ...AA Alkaline Batteries One year battery life with 10 second Automatic Shutoff Timer setting Displayed Units lb oz kg Capacity 45 lb x 0 2 oz 20 kg x 5 g Construction ABS Plastic Housing Options Carryin...

Страница 15: ...cale Scale will not come to zero when the ZERO button is pressed Make sure that the scale is stabilizing After pressing the ZERO button the scale should zero as soon as it becomes stable Weight readin...

Страница 16: ...an calibration is out of range Error is displayed after a span calibration attempt Refer to the Calibration Troubleshooting section for additional information If this problem persists contact Doran Te...

Страница 17: ...or indirect damages to equipment or personnel Nor shall Doran Scales Inc be liable for damages to equipment or for personal injury caused by misuse overload accidental damage alteration improper inst...

Страница 18: ......

Отзывы: