c. When the input is from the "A" terminal, excessive current will blow the fuse which must then be replaced.

Fuse value is 0.5A.

d. The 20A range is not protected by a fuse and has a maximum 10A continuous current rating. The

20A range measurement time should not exceed 15 seconds.

5.3 RESISTANCE MEASUREMENT

1. Connect black test lead to "COM" terminal and red test lead to the "V/OHM" input terminal.

2. Set the function range switch to the OHM range.

3. Connect the test leads across the resistance under measurement and read the display value.

Note:

a. The polarity of the red test lead is "+".

b. When the input is not connected, i.e at open circuit, the figure "1" will be displayed for the over

range condition.

c. If the resistance value being measured exceeds the maximum value of the range selected an over

range indication "1" will be displayed and function range switch must be set to a higher range.

d. 200MΩ range has a 10 digits (1MΩ) constant. This figure will appear in short circuit status and it should be

subtracted from any measurement result on this range. For instance: when measuring a100MΩ resistor a figure

of 101.0 will be shown in the display. A value of 1MΩ should be subtracted from the indicated value.

5.4 CAPACITANCE MEASUREMENT

1. Set the function range switch to the "Cx" position. Before connecting the capacitor, wait until the display

automatically zeros.

2. Connect the test capacitor to the "Cx" input socket (not test leads) and read the display value.

3. When using the probes for testing capacitors , insert the probes into the “CX” hole(Marked

with “mA & COM“) and connect test lead tips to the capacitor, then read the display value.

Note:

The tested capacitor should be discharged before the testing procedure. Never apply voltage to the "Cx"

input socket, or serious damage may result.

5.5 DIODE & CONTINUITY TEST

1. Set the function range switch to the

position.

2. Connect the black test lead to "COM" terminal and red test lead to V/OHM input terminal; (Note: the polarity

of the red test lead is "+").

3. This range has an "AUDIBLE CONTINUITY TEST" function. Built-in buzzer sounds if the resistance

between the two probes is less than 30±10Ω.

4. Connect the test leads across the diode and read the display value.

Note:

a. When the input is not connected, i.e. at open circuit, the figure "1" will be displayed

b. Test condition: Forward DC current approx 1mA. Reversed DC voltage approx. 2.8V

c. The meter displays the forward voltage drop and displays figure "1" for overload when the diode

is reversed.

5.6 TRANSISTOR hFE TEST

1. Set the function range switch to the "hFE" position.

2. Determine whether the transistor is "NPN" or "PNP" type.

3. Correctly insert transistor into E.B.C connector.

4. Displayed reading is approx. transistor hFE value.

Note:

Test condition: Base current approx 10uA. Vce approx.2.8V

6. BATTERY AND FUSE REPLACEMENT

1. Battery and fuse replacement should only be performed after the test leads have been disconnected and power

turned off.

2. Loosen screws with suitable screwdriver and remove case bottom

3. The meter is power by a single 9V battery. Snap the battery connector leads to the terminals of a new

battery and reinsert the battery into the case top. Arrange the battery leads so that they will not be pinched

between the case bottom and case top.

4. The meter is protected by a fast fuse 0.5A/250V. Fuse dimensions are 5mmФ x 20mm.it is soldered on the

PCB on some versions.

5. Replace the case bottom and reinstall the three screws. Never operate the meter unless the case bottom is

in place.

Содержание UA9205N

Страница 1: ...yed when the battery voltage drops below the operating voltage Full range over load protection Capacitance measurement Auto zeroing Auto Power Off Automatic cut off approx 15 minutes after the power on It needs to be switched off and on again to continue operation Operating temperature 0 C 40 C 0 75 R H Storage temperature 10 C 50 C 0 75 R H Power Single standard 9V battery Dimensions 191mm L x 95...

Страница 2: ...ccuracy 2mA 1 3 20mA 200mA 1 8 3 20A 2 5 Measuring voltage drop 200mV 3 4 ACA Range Accuracy 2mA 0 8 1 20mA 200mA 1 2 1 20A 2 5 Measuring voltage drop 200mV Frequency range 40 400Hz 3 5 CAPACITANCE Range Accuracy 2nF 20nF 200nF 2 5 3 2uF 20uF 3 6 OHM Range 200Ω 2K Ω 20K Ω 200K Ω 2M Ω 20M Ω 200M Ω Accuracy 0 8 3 0 8 1 1 2 5 1 ...

Страница 3: ... Do not use or store the instrument in conditions of high temperature and high humidity 5 METHOD OF MEASUREMENT 5 1 DCV ACV MEASUREMENT 1 Set the Function range switch at the required position 2 Connect black test lead to COM terminal and red test lead to the V OHM input terminal 3 Connect test leads to measuring point and read the display value The polarity of the red lead connection will be indi...

Страница 4: ...ested capacitor should be discharged before the testing procedure Never apply voltage to the Cx input socket or serious damage may result 5 5 DIODE CONTINUITY TEST 1 Set the function range switch to the position 2 Connect the black test lead to COM terminal and red test lead to V OHM input terminal Note the polarity of the red test lead is 3 This range has an AUDIBLE CONTINUITY TEST function Built...

Отзывы: