background image

4. PRECAUTIONS AND PREPARATIONS FOR MEASUREMENT

1. Ensure the battery is correctly placed in the battery compartment.

2. Do not exceed the input limits shown below:

Function Range

Input Terminals

Maximum Input

DCV 200mV

250VDC

ACV 200mV

V/OHM COM

250VAC

DCV 2~1000V

1000 VDC

ACV 2~750V

750VAC

OHM

V/OHM COM

250V DC/AC

Diode

V/OHM COM

DCA 200mA

200mA DC/AC

ACA 200mA

A COM

DCA2A

2A DC/AC

ACA2A
DCA 20A

20A COM

20A DC/AC

3. Inspect the test leads for damaged insulation or exposed metal. Check Test lead continuity. Damaged

leads should be replaced.

4. Select the proper function and range for your measurement

5. Check the input terminal position for red test lead depending on measurement ranges.

6. Either one of the test leads should be removed from the circuit under test when changing the test ranges.

7. To avoid electrical shock or damage to the meter Do not apply more then 500V between any terminal and

earth ground.

8. To avoid electrical shock, use caution when working above 60VDC or 25VAC rms. Such voltage poses a

shock hazard.

9. When measurement is completed switch off the power. Remove the battery when not used for

extended periods to avoid leakage problems.

10. Do not tamper with the circuitry to avoid damage.

11. Do not use or store the instrument in conditions of high temperature and high humidity.

5. METHOD OF MEASUREMENT

5.1 DCV & ACV MEASUREMENT

1. Set the Function range switch at the required position.

2. Connect black test lead to "COM" terminal and red test lead to the "V/OHM" input terminal.

3. Connect test leads to measuring point and read the display value. The polarity of the red lead connection

will be indicated at the same time as the voltage.

Note:

a. If the voltage to be tested is unknown beforehand, set the Function range switch to the highest range

and work down.

b. When only the figure "1" is displayed over range is being indicated and the function range switch has to

be set to a higher range.

c. Never try to measure voltage above 1000V! Although indication is possible, there is danger of damaging

the internal circuitry.

5.2 DCA & ACA MEASUREMENT

1. Connect the black test lead to the "COM" terminal and the red test lead to "A" terminal for a maximum of 0.5A

2. Set the function range switch at the required position.

3. Connect test leads to measuring points and read the display value. The polarity at the red test lead

connection will be indicated at the same time as the current.

Note:

a. If the current range is unknown beforehand, set the function range switch to the highest range and

work down.

b. When only the figure "1" is displayed, over range is being indicated and the function range switch has

be set to a higher range.

Содержание UA9205N

Страница 1: ...yed when the battery voltage drops below the operating voltage Full range over load protection Capacitance measurement Auto zeroing Auto Power Off Automatic cut off approx 15 minutes after the power on It needs to be switched off and on again to continue operation Operating temperature 0 C 40 C 0 75 R H Storage temperature 10 C 50 C 0 75 R H Power Single standard 9V battery Dimensions 191mm L x 95...

Страница 2: ...ccuracy 2mA 1 3 20mA 200mA 1 8 3 20A 2 5 Measuring voltage drop 200mV 3 4 ACA Range Accuracy 2mA 0 8 1 20mA 200mA 1 2 1 20A 2 5 Measuring voltage drop 200mV Frequency range 40 400Hz 3 5 CAPACITANCE Range Accuracy 2nF 20nF 200nF 2 5 3 2uF 20uF 3 6 OHM Range 200Ω 2K Ω 20K Ω 200K Ω 2M Ω 20M Ω 200M Ω Accuracy 0 8 3 0 8 1 1 2 5 1 ...

Страница 3: ... Do not use or store the instrument in conditions of high temperature and high humidity 5 METHOD OF MEASUREMENT 5 1 DCV ACV MEASUREMENT 1 Set the Function range switch at the required position 2 Connect black test lead to COM terminal and red test lead to the V OHM input terminal 3 Connect test leads to measuring point and read the display value The polarity of the red lead connection will be indi...

Страница 4: ...ested capacitor should be discharged before the testing procedure Never apply voltage to the Cx input socket or serious damage may result 5 5 DIODE CONTINUITY TEST 1 Set the function range switch to the position 2 Connect the black test lead to COM terminal and red test lead to V OHM input terminal Note the polarity of the red test lead is 3 This range has an AUDIBLE CONTINUITY TEST function Built...

Отзывы: