CM170407AMA
NOTE
the AoA channel can be controlled by the test profile, but it is not recorded into the
profile results.
In addition to the test points, a test profile also contains:
•
limiting values for the controlled quantities,
•
measure units selection,
•
altitude offset to be applied.
11.1.2 MEMORY ORGANIZATION
The non-volatile memory of the MPS49 contains 30 locations for storing the profiles. Every
time a profile is saved into one memory location, it overwrites the previously saved profile in
the same location.
The same thing applies to the profile results. The MPS49 contains 300 profile results.
11.2 ENTERING TEST PROFILES
A test profile can be changed or created by touching the “Edit” button on the Profiles Menu.
The MPS49 will show the Profile Selection Menu for selecting the memory location. If the
Operator selects an empty memory location, a new profile will be created.
After pressing the Select button, the display will show the Profile Editing Menu (FIG 22).
This menu allows the Operator to set:
•
name of the profile,
•
controlled units (aeronautical or pressures),
•
AoA control (2 or 3 lines operation),
•
number of steps,
•
altitude offset (only when controlling aeronautical units),
•
limiting values.
NOTE
: if the profile name begins with “leak”, the profile will be listed among the automatic
leak tests. Refer to paragraph 7.2.3 for more information.
FIG 22 THE PROFILE EDITING MENU
Touch the Steps button, to open the Profile Step Selection screen (FIG 23). For each step,
the MPS49 displays:
•
the altitude/Ps target;
•
the airspeed/Pt/Qc target;
D.Marchiori Copyright
43