user-defined limits for all controlled values and rates;
•
protection against Pt < Ps;
•
protection against excessive leaks in MEASURE mode.
In case of AC power loss, the MPS49 automatically switches to battery operation.
If the battery is flat or not available, the static, pitot and AoA lines remain sealed.
When power is restored, the MPS49 senses the current pressure values and begins
controlling again.
In case of emergency, the manual venting valves enable the Operator to safely bring all
lines back to ambient pressure.
2.2.12 EXTERNAL SUPPLY PORTS
The MPS49 vacuum source is connected to a pneumatic port on the front panel. The port
supports the use of DMA adaptor kits with suction cups.
2.2.13 OPTIONAL MULTIPLE ISOLATOR FUNCTION
The MPS49 can be provided with up to four Ps, four Pt and four AoA connections.
The Multiple Isolator function enables the Operator to individually seal each port, to allow
advanced leak checking procedures and generation of different pressure levels.
2.2.14 ADDITIONAL CONNECTIONS
The MPS49 comes equipped with an RS232 port, that enables communication with an
external PC for remote control. Refer to SECTION 12 for more information. The MPSRC
also connects to the RS232 port.
One type A USB port is also provided on the front panel. The MPS49 enables transferring
profiles and profile results to and from USB pen drives.
An optional Encoding Altimeter connector enables the MPS49 to receive gray code input.
Additional data communication ports are available as options:
•
IEEE 488 (GPIB);
•
ARINC 429;
•
Ethernet.
Finally, other connectivity options, such as wireless LAN and Bluetooth, are available as
external dongles.