You can use the
and
buttons to adjust the value displayed to match the reading of an external measurement
device (in his case, most likely a digital multimeter). Pressing
will revert the value displayed to the original
measurement.
If this is the last point to calibrate, pressing
will confirm and store the calibration values and leave calibration
mode; the display will be updated in
∼
2 seconds and you will see a corrected reading. If there are more points to
calibrate, pressing
saves the current value and lets you input another calibration point.
You can cancel calibration at any time by pressing
.
You can reset calibration to default by pressing
when not in calibration submode. You will be asked to confirm
this by pressing
, or decline by pressing
.
5.3
Network mode
Network mode
In Network mode, the
and
keys allow to select the network parameter. The selected parameter name is
displayed together with its current value.
Network configuration submode
The following parameters are present:
Parameter
Type
Default/sample value
HTTP port
TCP port
∗
80
HTTPS port
TCP port
∗
443
SSH port
TCP port
22
SSH enabled
Yes/No
No
Syslog server
IP address
0.0.0.0 (none)
LAN protocol
Static/DHCP
Static
LAN IP
IP address
192.168.0.1
LAN netmask
Netmask
255.255.255.0
LAN gateway
IP address
0.0.0.0 (none)
LAN DNS server
IP address
∗
0.0.0.0 (none)
LAN MAC
MAC
7C:E1:FF:..:..:..
WiFi enabled
Yes/No
Yes
WiFi SSID
String
DLI....
WiFi mode
Access Point/Station
Access point
WiFi encryption
None/WPA/WPA2
WPA
DLI LPC9 User’s Guide: 1.7.24.0
Содержание LPC9
Страница 1: ...DLI LPC9 User s Guide 1 7 24 0 ...
Страница 81: ......