![DLAB Laboratory Instruments D3024 Скачать руководство пользователя страница 13](http://html1.mh-extra.com/html/dlab-laboratory-instruments/d3024/d3024_user-manual_2512852013.webp)
High Speed (Refrigerated) Micro-Centrifuge
12
Figure 6-3 The main interface (D3024R)
Main interface of D3024R model is as figure 6-3. The speed is set to be 15000rpm, temperature of
centrifugal chamber is 4 ,
℃
and the running time is 12 minutes. When speed symbol
is rotating,
indicating the centrifuge is running, the rotation is faster, the speed is higher. When temperature symbol
is blinking, indicating refrigeration is running,
no blinking means refrigeration stop working.
Time symbol
displays the ratio of working to time setting. The total time setting is divided into 10
scales.
7. Rotor Preparation
7.1 Prepare the sample
7.2 Inject the sample into tubes.
CAUTION: Do not inject overloading samples into the centrifuge which will be
caused leaking.
Do not exceed the actual capacity allowed in the user manual.
7.3 Keep the tubes balance
Although the centrifuge can accept sample balancing by eye, we recommend that you use this
centrifuge in a well-balanced condition to extend its life expectancy.
Never intentionally run the centrifuge under unbalanced condition even though the allowable
imbalance is not exceeded.
7.4 Inspect the rotor
15000 4
℃
12
min
rpm