DLAB Laboratory Instruments D3024 Скачать руководство пользователя страница 1

 

 

 

 

 

 

 

 

 

 

 

 

 

 

     

 

 

 

 

 

 

D3024 & D3024R   

High Speed (Refrigerated) Micro-Centrifuge 

 

 

 

 

 

 

 

 

 

 

Dragon Laboratory Instruments Limited

   

www.dragon-lab.com 

 
 

 
 
 
 
 

Before using this centrifuge, please carefully read this user manual for its efficient operation and 
your safety. 

 

Содержание D3024

Страница 1: ...R High Speed Refrigerated Micro Centrifuge Dragon Laboratory Instruments Limited www dragon lab com Before using this centrifuge please carefully read this user manual for its efficient operation and...

Страница 2: ...reparation 12 7 1 Prepare the sample 12 7 2 Inject the sample into the tubes 12 7 3 Keep the tubes balance 12 7 4 Inspect the rotor 12 7 5 Symmetrically load centrifuge tubes in rotor 13 8 Operation 1...

Страница 3: ...ntrifuge 26 14 2 Guarantee of the rotor 26 After sales service 26 Copyright No part of this manual may be reproduced or transmitted without prior written permission of DRAGONLAB We can not be responsi...

Страница 4: ...WARNING Personal Danger Warning notes indicate any condition or practice which if not strictly observed could result in personal injury or possible death CAUTION Possible damage to instrument Caution...

Страница 5: ...advance and then notice the service center involved in the details of the particular materials z Do not handle the power cord or turn on or off the POWER switch with wet hands to void electrical shoc...

Страница 6: ...e allowable minimum speed of the rotor kits rotor bucket or the other adapter Make sure to run it below the allowable minimum speed 3 Do not exceed the allowable imbalance 4 Use the rotor and tubes wi...

Страница 7: ...z 60Hz 5A D3024R Single phase 220V 240V 50Hz 60Hz 10A 110V 120V 50Hz 60Hz 10A Dimensions mm D3024 L 280 W 364 H 266 D3024R L 338 W 580 H 324 Weight D3024 12kg D3024R 30kg Additional features Speed RCF...

Страница 8: ...to ensure your safety and the optimum performance Before moving the centrifuge the rotor must be removed WARNING Incorrect power supply may damage centrifuge Make sure the power source conforms to the...

Страница 9: ...grounded properly 1 This centrifuge is equipped with a 3P flat plug grounding can be done just by plugging the 3P plug into the outlet 2 An over 10A outlet which has a good ground is required and meet...

Страница 10: ...rated Micro Centrifuge 9 Figure 5 2 D3024 Rear view of centrifuge D3024R Figure 5 3 D3024R Front view of centrifuge Outlet Seal ring of door Operation panel Door lock hook POWER switch Cover Outlet Ai...

Страница 11: ...ed Refrigerated Micro Centrifuge 10 Figure 5 4 D3024R Rear view of centrifuge 6 Operation panel Air vents View Door 1 2 4 3 Speed display area Temperature display area Time display area 5 15000 25 12...

Страница 12: ...the button during centrifugation 5 Parameter button Clockwise rotate to increase parameter values counter clockwise rotate to decrease parameter values Press down the button shift between speed and RC...

Страница 13: ...ol displays the ratio of working to time setting The total time setting is divided into 10 scales 7 Rotor Preparation 7 1 Prepare the sample 7 2 Inject the sample into tubes CAUTION Do not inject over...

Страница 14: ...the rotor may be moved off while rotating and cause damage of the centrifuge or rotor z Firmly tighten the rotor door with rotor 8 Operation 8 1 Normal operation Turn on the POWER switch centrifuge w...

Страница 15: ...figure indicates the instrument has accumulated running time 312 hours 56 minutes and 45 seconds and then the instrument displays the last running parameters Figure 8 3 Last running interface Speed 15...

Страница 16: ...tor and ensure is in position Close the door and then start running The method of removing rotor is as same as the above mentioned by turning the screw counterclockwise 8 1 2 Set the operation paramet...

Страница 17: ...3 Start the operation 1 Press button to start running The door should be locked before rotor starts rotating It will start timing once the rotor starts running the screen displays remaining run time 2...

Страница 18: ...eration 1 Turn on the POWER switch 2 Set a RCF value CAUTION Do not exceed the allowable maximum RCF value of the rotor and adapters Press the select button choose speed unit g the speed symbol will f...

Страница 19: ...t with a cloth or sponge moistened with a neutral detergent solution Sterilize the centrifuge by wiping with a cloth moistened with 70 ethanol solution 2 Rotor chamber Once the rotor chamber is dirty...

Страница 20: ...below is the centrifuge consumables It is recommended to replace consumables according the table below No Consumables Replacement conditions 1 Seal ring of door Crack 2 Rubber block of temperature sen...

Страница 21: ...it is not available to operate the alarm code will be displayed on the time display screen and the operator can know the malfunction caused by the alarm code WARNING Do not open the door when the roto...

Страница 22: ...lly once the operation finished 3 Press button once the rotor stops door can be released 10 2 2 In the case of power outage If the door cannot be opened due to the power outage open the door as follow...

Страница 23: ...tor structure 11 1 2 Available rotors and adapters All rotors are used for bio safe when the rotor lid was tightened with the rotor centrifuge tubes will be enclosed into rotor to ensure the sample do...

Страница 24: ...ly a thin coat of silicon grease on the rotor surface after drying 4 It is necessary for a regular maintenance for rotor recommend to clean it each 3 months to keep cleaning of tube hole and shaft hol...

Страница 25: ...9 11 2 3 Autoclaving PA PC and PP tubes PA begins softening at about 120 PC and PP at about 130 Autoclave PA tubes at 115 0 7kg cm2 for 30 minutes and PC and PP tubes at 121 0 1kg cm2 for 20 minutes...

Страница 26: ...High Speed Micro Centrifuge with AS24 2 rotor kits Euro plug 220V 50Hz 60Hz 912315120000 D3024 High Speed Micro Centrifuge with AS24 2 rotor kits UK plug 220V 50Hz 60Hz 922015130000 D3024R High Speed...

Страница 27: ...even if within the guarantee period expires 1 Failures caused by incorrect installation 2 Failures caused by rough or improper handling 3 Failures caused by conveyance or relocation after installation...

Страница 28: ...fuge 27 E mail info dragon lab com Service service dragon lab com Website Http www dragon lab com US office DRAGONLAB LLC 500 Four Rod Road Suite 122 Berlin CT 06037 USA Tel 1 860 828 5289 Fax 1 860 8...