Document no. 477-2010A page 5 of 5 

Remote Control. 

 
All the functions of the front buttons are available through the 9pole D connector remote control connector.  
Previous users of the NTP477-series instrument will find the same possibilities in this 477-100 series although the 
needed external connections have changed. 
In addition to the remote control of operating functions (i.e. ”Disp. Peak”), also a number of operating parameters (i.e. 
ref. level) can be programmed externally. 
Restricting the number of control inputs to just four prohibits the use of simple external on/off connections. Instead 
specific resistance values must be used in series with the external switches/connections. The switches “common side” 
must all be connected to GND (Pin 4).  
 

Remote Control input 1 (RC#1) 

This input facilitates an alternative way of controlling the PPM reference level. If this input is left open the reference 
level is set as described under 

Operating Parameter Settings

 .  If however a resistor is connected to 0V the preset 

reference level will be overruled. The resistor value required to obtain a given reference level can be found in table 1. It 
may be necessary to use resistors in parallel or series to obtain a specific value to /-2% 
 
Table 1. 
 

Reference

Resistor Reference Resistor 

level 

value (ohms)

level 

value (ohms)

  

  

  

  

Reserved

-16 dB FS 

10625 

0 dB FS 

313 

-17 dB FS 

12000 

-1 dB FS 

645 

-18 dB FS 

13571 

-2 dB FS 

1000 

-19 dB FS 

15385 

-3 dB FS 

1379 

-20 dB FS 

17500 

-4 dB FS 

1786 

-21 dB FS 

20000 

-5 dB FS 

2222 

-22 dB FS 

23000 

-6 dB FS 

2692 

-23 dB FS 

26667 

-7 dB FS 

3200 

-24 dB FS 

31250 

-8 dB FS 

3750 

-25 dB FS 

37143 

-9 dB FS 

4348 

-26 dB FS 

45000 

-10 dB FS

5000 

-27 dB FS 

56000 

-11dB FS

5714 

-28 dB FS 

72500 

-12 dB FS

6500 

-29 dB FS 

100000 

-13 dB FS

7368 

-30 dB FS 

155000 

-14 dB FS

8333 

-31 dB FS 

320000 

-15 dB FS

9412 

Preset level 

N.C. 

 
 
 
 
 
Remote Control input 2 (RC#2) 

This input is reserved for controlling the reference level loudness mode. (Not implemented in this version) 
 
 

Remote Control input  (RC#3) 

is reserved for future use. 

 
 

Содержание 477-200A

Страница 1: ...20030929 Document no 477 2010A page 1 of 1 Digital Peak Programme Meter 477 200A Users Manual ...

Страница 2: ...The spot responds to the fastest peaks on the signal 0 integration time The calibration of the spot reading is such that the mark 0 or the start of overload range equals digital full scale level This means that the distance between the spot indication and the mark 0 on the shown scale represents the true head room before digital clipping This means that even with an analogue type scale and 10ms in...

Страница 3: ... reference level in 1 dB steps Scale max equals digital FS 3 Press 4 to increase reference level in 1 dB steps Scale max equals digital FS 4 Press 1 to save settings and resume to normal operation Or press 2 to jump to next adjustment point 2 Activating deactivating display ballistics 1 Press 1 3 and hold for more than 1 6 s The display will be lit in both channels with markings pr 1 dB 2 Press 2 ...

Страница 4: ...he two connectors should be in the 110 Ώ position Power connector Power is supplied through the coaxial low voltage socket A mating connector should have an outher diameter of 5 5mm and a hole diameter of 2 1mm Either a 24V dc or a 18V ac sypply should be used Remote Control Remote control and additional power connection is accessible through the 9pole male D sub connector on the rear side Pin con...

Страница 5: ...ference level is set as described under Operating Parameter Settings If however a resistor is connected to 0V the preset reference level will be overruled The resistor value required to obtain a given reference level can be found in table 1 It may be necessary to use resistors in parallel or series to obtain a specific value to within 2 Table 1 Reference Resistor Reference Resistor level value ohm...

Страница 6: ...e 0 to 31 dB below max digital code Overload indication the bar intensity is increased within overload range Low frequency cut off DC blocking Cut off frequency 0 3 Hz Secondary reading spot Integration time zero Return time 1 7 s 0 3 s 0 to 20 dB IEC 268 10 1991 03 Scale max Scale max equals the lower limit of intensified bar range Reference level Scale max corresponds to max digital code level P...

Отзывы: