Adjustment ranges (min./max.)
Femur range:
approx. 31 to 49 cm
Tibia range:
approx. 25 to 57 cm
Accuracy of measured values
Goniometer in the measuring range: from − 10° to + 120°
Accuracy:
+/− 2°
Weight:
11 kg
Materials used:
ABS, POM, PUR, PA, FR 4,
aluminum, stainless steel, brass
MDD:
class IIa
Standards compliance:
93/42/EEC
IEC 60601-1
IEC 60601-1-6
IEC 60601-1-11 (S# > 20,000)
IEC 62366
IEC 62304
EN ISO 14971
ANSI/UL 60601-1 (S# < 20,000)
CAN/CSA C22.2 No. 601.1-M90 (S# < 20,000)
ANSI AAMI ES 60601-1 1st Edition (S# > 20,000)
CAN CSA 22.2 No. 60601-1-08 (S# > 20,000)
EMC
(electromagnetic compatibility)
IEC 60601-1-2
Manufactured in compliance with:
EN ISO 13485
Ambient conditions (storage, transport)
Temperature:
Relative humidity:
Atmospheric pressure:
− 25 °C to + 70 °C
at 70 °C to 93 % no condensation
500 hPa to 1060 hPa
Ambient conditions (operation)
Temperature:
Relative humidity:
Atmospheric pressure:
+5 °C to +40 °C
15 % to 93 %
700 hPa to 1060 hPa